FDA RECALL RECORD · Z-0414-2026

Xpert SA Nasal Complete Recall

Xpert SA Nasal Complete Catalog Number: GXSACOMP-10 UDI-DI code: 07332940000158 Xpert SA Nasal Complete kit; 10 test kit: A collection of reagents and other associated materials intended to be used for the direct qualitative and/or quantitative detection of nucleic acid from methicillin-resistant and methicillin-susceptible Staphylococcus aureus (MRSA/MSSA) bacteria present in a clinical specimen, using a nucleic acid technique

FDA recall Class II

Ongoing

FDA lists this Class II recall for Xpert SA Nasal Complete. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0414-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 6, 2025
FDA report date
Nov 5, 2025
Quantity in commerce
4,132 kits
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Xpert SA Nasal Complete Catalog Number: GXSACOMP-10 UDI-DI code: 07332940000158 Xpert SA Nasal Complete kit; 10 test kit: A collection of reagents and other associated materials intended to be used for the direct qualitative and/or quantitative detection of nucleic acid from methicillin-resistant and methicillin-susceptible Staphylococcus aureus (MRSA/MSSA) bacteria present in a clinical specimen, using a nucleic acid technique

Affected model & UDI checker

Search the exact identifiers published for Z-0414-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
GXSACOMP-10Xpert SA Nasal CompleteCheck source noticeNot specified
Not specifiedXpert SA Nasal CompleteCheck source notice07332940000158
FDA code information

Catalog Number: GXSACOMP-10 UDI-DI code: 07332940000158 Lot Numbers: 27815 27816 Batch Numbers: 1001424179 1001424181

View the original FDA recall notice

Why was this product recalled?

As a result of Post-Market half-time stability testing, false negative test result was observed for a positive sample type.

What actions did Cepheid take?

On 08/21/2025, the firm sent via FedEx an "Urgent Medical Device Recall" to customers to inform customers that Cepheid conducted a routine post-market stability study for GXSACOMP-10 kit batch 1001424179 lot 27815, which failed to meet the half-time stability criteria set for this product. Preliminary investigation indicates that kit batch 1001424181 lot 27816 of the same product may also be impacted. Customers are instructed to: 1. Dispose of GXSACOMP-10 lots 27815 and/or 27816 you have in inventory. Cepheid will provide replacement product. The replacement product should ship within 5 business days following receipt of the completed response form. 2. If you have previously received or receive a negative test result with GXSACOMP-10 lots 27815 or 27816, and the patient shows clinical signs/symptoms of SA or MRSA, re-test with a cartridge from a different lot or use an alternative test/method for SA or MRSA, if available. 3. With all questionable positive or questionable negative results, please continue to report to Cepheid Technical Support for investigation. 4. Cepheid asks that you acknowledge receipt of this Urgent Medical Device Recall by completing the attached Customer Response Form and either emailing the completed form to [email protected] or faxing it to +1 (408) 716-3143. Cepheid will not send a reply or confirmation of received response forms.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.