Brand Name: Sentinel V11 Product Name: Sentinel V11 Recall
Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Sentinel V11 Product Name: Sentinel V11. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0416-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 16, 2025
- FDA report date
- Nov 12, 2025
- Quantity in commerce
- 142
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.
Affected model & UDI checker
Search the exact identifiers published for Z-0416-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 92810 | Brand Name: Sentinel V11 Product Name: Sentinel V11 | Check source notice | Not specified |
| Not specified | Brand Name: Sentinel V11 Product Name: Sentinel V11 | Check source notice | 10841522125751 |
FDA code information
Catalog Number: 92810 UDI-DI code; (01)10841522125751 Software version: 11.6.0
Why was this product recalled?
Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
What actions did Spacelabs Healthcare, Ltd. take?
On 09/15/2025, the firm sent via email an "URGENT MEDICAL DEVICE CORRECTION" Letter to customer informing them of two issues associated with the Sentinel Software Version 11.6.0. The issues are: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record. Customers are instructed to: 1. Print and Post Notification: Please print the related Urgent User Notification (Appendix A) in all location(s) where Sentinel 11.6.0 is used to configure ABP and/or Ambulatory ECG (Holter) devices. 2. Notify all relevant employees within their organization, including satellites, of this recall 3. Confirm the above action has been taken: Go to the QR code below (Appendix B) and confirm that they have read, understood, and posted the user notification at their workstations. 4. Contact Customer Support team at [email protected] with any questions regarding this identified issue. A fix for these issues is being developed, and is anticipated to be available by the end of September 2025. A Spacelabs Healthcare representative will notify customers once the fix is available and will work with customers to schedule the software update. Note, rolling back Sentinel 11.6.0 to a previous version is not a valid option.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0416-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology