Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0418-2026

Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing Recall

Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing, Model/Catalog Number: 110031012; semi constrained hip prosthesis

FDA recall Class II

Ongoing

FDA lists this Class II recall for Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0418-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 26, 2025
FDA report date
Nov 12, 2025
Quantity in commerce
72 units
Distribution
See key details
FDA cause
Labeling mix-ups

What is being recalled?

Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing, Model/Catalog Number: 110031012; semi constrained hip prosthesis

Affected model & UDI checker

Search the exact identifiers published for Z-0418-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
110031012Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, BearingCheck source noticeNot specified
Not specifiedZimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, BearingCheck source notice00889024572706
FDA code information

UDI (01)00889024572706(17)300406(10)67160447; Lot Number 67160447

View the original FDA recall notice

Why was this product recalled?

Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E¿ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.

What actions did Zimmer, Inc. take?

Zimmer Biomet issued an URGENT MEDICAL DEVICE recall notice to its consignees on 9/25/2025 via FedEx, 2nd Day Delivery. The notice explained the issue, potential risks, and requested the return of all affected devices in inventory. Risk Managers were notified that the sales representatives would be removing the affected products. Recommendation for patient monitoring was made to the implanting surgeons. For further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 a.m. and 5:00 p.m. EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, questions may be emailed to [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology