Brand Name: Klassic BiPolar Head Recall

Brand Name: Klassic BiPolar Head Model/Catalog Number: 3205.41.022 Klassic BiPolar Head, 41 mm OD, 22 mm Head; 3205.42.022 Klassic BiPolar Head, 42 mm OD, 22 mm Head; 3205.43.022 Klassic BiPolar Head, 43 mm OD, 22 mm Head; 3205.44.028 Klassic BiPolar Head, 44 mm OD, 28 mm Head; 3205.45.028 Klassic BiPolar Head, 45 mm OD, 28 mm Head; 3205.46.028 Klassic BiPolar Head, 46 mm OD, 28 mm Head; 3205.47.028 Klassic BiPolar Head, 47 mm OD, 28 mm Head; 3205.48.028 Klassic BiPolar Head, 48 mm OD, 28 mm Head; 3205.49.028 Klassic BiPolar Head, 49 mm OD, 28 mm Head; 3205.50.028 Klassic BiPolar Head, 50 mm OD, 28 mm Head; 3205.51.028 Klassic BiPolar Head, 51 mm OD, 28 mm Head; 3205.52.028 Klassic BiPolar Head, 52 mm OD, 28 mm Head; 3205.53.028 Klassic BiPolar Head, 53 mm OD, 28 mm Head; 3205.54.028 Klassic BiPolar Head, 54 mm OD, 28 mm Head; 3205.55.028 Klassic BiPolar Head, 55 mm OD, 28 mm Head; 3205.56.028 Klassic BiPolar Head, 56 mm OD, 28 mm Head; 3205.58.028 Klassic BiPolar Head, 58 mm OD, 28 mm Head; 3205.60.028 Klassic BiPolar Head, 60 mm OD, 28 mm Head

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Klassic BiPolar Head. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0432-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 23, 2025
FDA report date
Nov 12, 2025
Quantity in commerce
103 units
Distribution
See key details
FDA cause
Packaging

What is being recalled?

Brand Name: Klassic BiPolar Head Model/Catalog Number: 3205.41.022 Klassic BiPolar Head, 41 mm OD, 22 mm Head; 3205.42.022 Klassic BiPolar Head, 42 mm OD, 22 mm Head; 3205.43.022 Klassic BiPolar Head, 43 mm OD, 22 mm Head; 3205.44.028 Klassic BiPolar Head, 44 mm OD, 28 mm Head; 3205.45.028 Klassic BiPolar Head, 45 mm OD, 28 mm Head; 3205.46.028 Klassic BiPolar Head, 46 mm OD, 28 mm Head; 3205.47.028 Klassic BiPolar Head, 47 mm OD, 28 mm Head; 3205.48.028 Klassic BiPolar Head, 48 mm OD, 28 mm Head; 3205.49.028 Klassic BiPolar Head, 49 mm OD, 28 mm Head; 3205.50.028 Klassic BiPolar Head, 50 mm OD, 28 mm Head; 3205.51.028 Klassic BiPolar Head, 51 mm OD, 28 mm Head; 3205.52.028 Klassic BiPolar Head, 52 mm OD, 28 mm Head; 3205.53.028 Klassic BiPolar Head, 53 mm OD, 28 mm Head; 3205.54.028 Klassic BiPolar Head, 54 mm OD, 28 mm Head; 3205.55.028 Klassic BiPolar Head, 55 mm OD, 28 mm Head; 3205.56.028 Klassic BiPolar Head, 56 mm OD, 28 mm Head; 3205.58.028 Klassic BiPolar Head, 58 mm OD, 28 mm Head; 3205.60.028 Klassic BiPolar Head, 60 mm OD, 28 mm Head

Affected model & UDI checker

Search the exact identifiers published for Z-0432-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
3205.41.022Brand Name: Klassic BiPolar HeadCheck source noticeNot specified
FDA code information

Part Number/UDI/Lot Number: 3205.41.022 00814703011931 3205.42.022 00814703011948 3205.43.022 00814703011955 3205.44.028 00814703011962 3205.45.028 00814703011979 3205.46.028 00814703011986 3205.47.028 00814703011993 3205.48.028 00814703012006 3205.49.028 00814703012013 3205.50.028 00814703012020 3205.51.028 00814703012037 3205.52.028 00814703012044 3205.53.028 00814703012051 3205.54.028 00814703012068 3205.55.028 00814703012075 3205.56.028 00814703012082 3205.58.028 00814703012105 3205.60.028 00814703012129

View the original FDA recall notice

Why was this product recalled?

Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Total Joint Orthopedics, Inc. take?

On September 5, 2025 Total Joint issued customer a Field action notification. On September 23, 2025, Total Joint issued a Urgent Medical Device Recall Notification. Total Joint ask consignees to take the following actions: 1. Stop implanting the affected product. 2. Locate and quarantine all affected inventory. 3. With preliminary info we initially sent a Market Withdrawal notification on 9/05/2025 to initiate the quarantining and return of all BiPolar Heads in your possession. 4. A list of all BiPolar Heads in your possession as of 9/05/2025 is shown within the letter, if you have already returned them we appreciate your swift response, if not please return these items immediately. 5. Please complete and return the attached acknowledgment form confirming receipt of this notice and detailing the quantities returned. If inventory discrepancies are identified, please note that in the Return Response Box. 6. Please contact Total Joint Orthopedics at (801)486-6070 with any questions or concerns.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.