Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0443-2026

Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550 Recall

Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550

FDA recall Class I

Ongoing

FDA lists this Class I recall for Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0443-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Oct 10, 2025
FDA report date
Nov 19, 2025
Quantity in commerce
96,406 total units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550

Affected model & UDI checker

Search the exact identifiers published for Z-0443-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedAnesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550Check source notice40889942469704
Not specifiedAnesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550Check source notice10889942469703
FDA code information

DYNJ10550 UDI-DI 40889942469704 (case), 10889942469703 (ea) Lot 25EBQ445

View the original FDA recall notice

Why was this product recalled?

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medline Industries, LP take?

Safety notice was provided to customers on 10/10/2025. Follow-up "URGENT MEDICAL DEVICE RECALL" notification dated 10/21/2025 was sent out on 10/21/2025. Notification was through email and first-class mail. Customers were instructed to Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product and destroy affected kits on hand. No products are to be returned to Medline. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-215-FG Recall Code: If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. An expanded "URGENT MEDICAL DEVICE RECALL" notification dated 10/24/2025 was provided to customers on 10/24/25, which included additional affected products. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.