Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0460-2026

Brand Name: Sonesta Product Name: 6210 Recall

Brand Name: Sonesta Product Name: 6210 Model/Catalog Number: 0120-LA120-03 Software Version: N/A Product Description: Sonesta 6210 - Procedure table designed for full fluoroscopic imaging of the urinary tract during video Urodynamics procedures. Component: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Sonesta Product Name: 6210. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0460-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 22, 2025
FDA report date
Nov 19, 2025
Quantity in commerce
10 tables
Distribution
See key details
FDA cause
Process control

What is being recalled?

Brand Name: Sonesta Product Name: 6210 Model/Catalog Number: 0120-LA120-03 Software Version: N/A Product Description: Sonesta 6210 - Procedure table designed for full fluoroscopic imaging of the urinary tract during video Urodynamics procedures. Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-0460-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
0120-LA120-03Brand Name: Sonesta Product Name: 6210Check source noticeNot specified
FDA code information

Lot Code: Primary DI Number: 07350006950018 Serial numbers of potentially affected units: 12-70791 12-70792 12-70793 12-70794 12-70795 12-70796 12-70797 12-70798 12-70799 12-70801

View the original FDA recall notice

Why was this product recalled?

Video fluoroscopy table may contain inadequate weld on backrest bracket resulting in backrest dropping.

What actions did Sonesta Medical AB take?

On September 8, 2025, Sonesta Medical AB issued a recall notification to affected consignees via E-Mail. Sonesta asked consignees to take the following actions: 1. Please review your table and confirm that the serial number on the product labeling matches with the number listed in the notification. 2. Please discontinue use of the table. It is recommended that you place a copy of this notice on the table to provide staff awareness of the issue. 3. Sonesta will contact you to schedule and perform a table inspection as soon as possible. 4. If the table inspection results are acceptable, Sonesta will provide you with evidence to continue to use the table. 5. If the table inspection results are unacceptable, Sonesta and Laborie will work with you to replace your table If you have any questions, contact the CEO via email: [email protected] or phone +468-502 572 80.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.