BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000; Recall

BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000;

FDA recall Class II

Ongoing

FDA lists this Class II recall for BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0479-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 2, 2023
FDA report date
Nov 19, 2025
Quantity in commerce
167 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000;

Affected model & UDI checker

Search the exact identifiers published for Z-0479-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1000BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000;Check source noticeNot specified
Not specifiedBEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000;Check source notice00860002987804
FDA code information

Model/Catalog: 1000; UDI-DI: 00860002987804; Lot number: 7008409;

View the original FDA recall notice

Why was this product recalled?

The field action was initiated in February of 2023 and was limited to the affected lot from sales force trunk stock only.

What actions did Miach Orthopaedics take?

On January 30, 2023 RE: BEAR Implant Hydration memo emails were sent to sales staff. If you observe an implant that hydrates too quickly (e.g., less than 30 seconds) and appears that it may not remain intact upon insertion into the knee, do not attempt to implant the device. We would recommend you always have a backup device available to address such circumstances. We plan to revise the Tips and Pearls document, ML-1027, to provide further instructions/details on how to properly hydrate the BEAR device. In the meantime, one way to avoid shortened hydration time is to consider hydrating the dense side of the implant first and continue toward the porous side . To date, all reports of this observation have involved Lot # 7008409. As a precaution and to ensure customer satisfaction, we are asking that you return all devices from this lot that are currently in your trunk stock (in your possession). Chris Brown will work with you to coordinate returns as well as replacing these trunk stock devices. Please contact me ([email protected]) or Arjun ([email protected]) if you have questions.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology