Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0482-2026

Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Recall

Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Model/Catalog Number: SD-400U-10 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0482-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 6, 2025
FDA report date
Nov 19, 2025
Quantity in commerce
48208 units
Distribution
See key details
FDA cause
Packaging

What is being recalled?

Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Model/Catalog Number: SD-400U-10 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.

Affected model & UDI checker

Search the exact identifiers published for Z-0482-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
SD-400U-10Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical SnareCheck source noticeNot specified
Not specifiedBrand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical SnareCheck source notice04953170408243
FDA code information

Model: SD-400U-10; Lot Numbers: 27V, 28V, 29V, 2XV, 2YV, 2ZV, 31V, 32V, 33V, 34V, 35V, 36V, 37V, 38V, 39V, 3XV, 3YV, 3ZV, 41V, 42V, 43V, 44V, 45V, 46V, 47V, 48V, 49V, 4XV, 4YV, 51V, 52V, 53V, 54V, 55V, 56V, 57V UDI: 04953170408243

View the original FDA recall notice

Why was this product recalled?

Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.

What actions did Olympus Corporation of the Americas take?

On October 6, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: Examine your inventory and quarantine any affected devices. Cease usage of the product immediately. Please contact customer service to request a return material authorization, a credit will be issued for any returned product. Please forward this notice to other users who may have the affected products if you have further distributed it. Olympus requests you to report any complaints, including holes in device packaging, to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology