FDA RECALL RECORD · Z-0487-2026

Affinity NT Oxygenator, Oxygenator with Balance Biosurface and Uncoated Cardiotomy/Venous Reservoir, Product Number 541B Recall

Affinity NT Oxygenator, Oxygenator with Balance Biosurface and Uncoated Cardiotomy/Venous Reservoir, Product Number 541B

FDA recall Class II

Ongoing

FDA lists this Class II recall for Affinity NT Oxygenator, Oxygenator with Balance Biosurface and Uncoated Cardiotomy/Venous Reservoir, Product Number 541B. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0487-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 12, 2025
FDA report date
Nov 19, 2025
Quantity in commerce
26 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Affinity NT Oxygenator, Oxygenator with Balance Biosurface and Uncoated Cardiotomy/Venous Reservoir, Product Number 541B

Affected model & UDI checker

Search the exact identifiers published for Z-0487-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedAffinity NT Oxygenator, Oxygenator with Balance Biosurface and Uncoated Cardiotomy/Venous Reservoir, Product Number 541BCheck source notice00763000187842
Not specifiedAffinity NT Oxygenator, Oxygenator with Balance Biosurface and Uncoated Cardiotomy/Venous Reservoir, Product Number 541BCheck source notice20763000187846
FDA code information

Lot Number: 231621003; Unit Serial Number: 5117755016; UDI-DI: 00763000187842, 20763000187846

View the original FDA recall notice

Why was this product recalled?

Medtronic identified an Affinity NT Oxygenator unit with decreased CO2 gas transfer rates that did not meet performance requirements.

What actions did MEDTRONIC PERFUSION SYSTEMS take?

An Urgent: Medical Device Recall notification letter dated September 2025 was sent to customers. Patient Recommendations: Patient previously supported with the impacted device face no additional risk from the issue described in this communication and should continue to be monitored per your practice s normal follow-up procedures. Customer Actions: Medtronic requests that you take the following actions: Review your inventory for listed lot number. If your inventory contains a product with Lot Number 231621003, please refer to the NT Affinity Oxygenator to identify the Serial Number. For guidance, consult Attachment A Identifying Affected Product. If you have the unit with Serial Number 5117755016 immediately quarantine and return it to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return and exchange or credit of this unit. Your Medtronic sales representative can assist you in the return of unit as necessary. Complete the enclosed Customer Confirmation Form and email to [email protected]. This form must be returned even if you do not have the affected product in your possession. Please maintain a copy of this communication in your records. Medtronic: Report complaints at 1-800-854-3570, option 4, or email to [email protected]. If you have any questions regarding this communication, please contact your Medtronic Field Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology