Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the X-ray tube and collimator to position the X-ray tube unit. This unit is used for radiographic examinations of whole body except mammography when combined with an X-ray high voltage generator, an X-ray tube unit, an X-ray collimator, and if necessary, an X-ray radiography stand, an X-ray radiography table, and digital radiography system. Recall
Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the X-ray tube and collimator to position the X-ray tube unit. This unit is used for radiographic examinations of whole body except mammography when combined with an X-ray high voltage generator, an X-ray tube unit, an X-ray collimator, and if necessary, an X-ray radiography stand, an X-ray radiography table, and digital radiography system.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the X-ray tube and collimator to position the X-ray tube unit. This unit is used for radiographic examinations of whole body except mammography when combined with an X-ray high voltage generator, an X-ray tube unit, an X-ray collimator, and if necessary, an X-ray radiography stand, an X-ray radiography table, and digital radiography system.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0499-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 16, 2025
- FDA report date
- Nov 19, 2025
- Quantity in commerce
- 1,766 devices
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the X-ray tube and collimator to position the X-ray tube unit. This unit is used for radiographic examinations of whole body except mammography when combined with an X-ray high voltage generator, an X-ray tube unit, an X-ray collimator, and if necessary, an X-ray radiography stand, an X-ray radiography table, and digital radiography system.
Affected model & UDI checker
Search the exact identifiers published for Z-0499-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| CH-200 | Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the X-ray tube and collimator to position the X-ray tube unit. This unit is used for radiographic examinations of whole body except mammography when combined with an X-ray high voltage generator, an X-ray tube unit, an X-ray collimator, and if necessary, an X-ray radiography stand, an X-ray radiography table, and digital radiography system. | Check source notice | Not specified |
| CH-200M | Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the X-ray tube and collimator to position the X-ray tube unit. This unit is used for radiographic examinations of whole body except mammography when combined with an X-ray high voltage generator, an X-ray tube unit, an X-ray collimator, and if necessary, an X-ray radiography stand, an X-ray radiography table, and digital radiography system. | Check source notice | Not specified |
| Not specified | Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the X-ray tube and collimator to position the X-ray tube unit. This unit is used for radiographic examinations of whole body except mammography when combined with an X-ray high voltage generator, an X-ray tube unit, an X-ray collimator, and if necessary, an X-ray radiography stand, an X-ray radiography table, and digital radiography system. | Check source notice | 04540217014484 |
FDA code information
Model Numbers: CH-200. UDI code: 04540217014484. Serial Numbers: 3ZC5C2A84014, 3ZC5C2A85003, 3ZC5C2A77013, 3ZC5C2A86013, 3ZC5C2A75012, 3ZC5C2A76002, 3ZC5C2A76016, 3ZC5C2A76026, 3ZC5C2A79003, 3ZC5C2A84008, 3ZC5C2A73001, 3ZC5C2A87008, 3ZC5C2A7A001, 3ZC5C2A76018, 3ZC5C2A79001, 3ZC5C2A7A002, 3ZC5C2A7B001, 3ZC5C2A85015, 3ZC5C2A81001, 3ZC5C2A86003, 3ZC5C2A7C007, 3ZC5C2A84002, 3ZC5C2A7B019, 3ZC5C2A77010, 3ZC5C2A73006, 3ZC5C2A74004, 3ZC5C2A75002, 3ZC5C2A75005, 3ZC5C2A75006, 3ZC5C2A75007, 3ZC5C2A76010, 3ZC5C2A76013, 3ZC5C2A76022, 3ZC5C2A77011, 3ZC5C2A77012, 3ZC5C2A77014, 3ZC5C2A77023, 3ZC5C2A78002, 3ZC5C2A78003, 3ZC5C2A79002, 3ZC5C2A7C006, 3ZC5C2A7C016, 3ZC5C2A82004, 3ZC5C2A83002, 3ZC5C2A85010, 3ZC5C2A85016, 3ZC5C2A85018, 3ZC5C2A86002, 3ZC5C2A86009, 3ZC5C2A86011, 3ZC5C2A87010, 3ZC5C2A88003, 3ZC5C2A88005, 3ZC5C2A81002, 3ZC5C2A85017, 3ZC5C2A7B002, 3ZC5C2A86004, 3ZC5C2A85005, 3ZC5C2A71003, 3ZC5C2A86010, 3ZC5C2A81004, 3ZC5C2A73012, 3ZC5C2A73013, 3ZC5C2A73014, 3ZC5C2A76006, 3ZC5C2A76019, 3ZC5C2A76027, 3ZC5C2A77003, 3ZC5C2A7B015, 3ZC5C2A85019, 3ZC5C2A88008, 3ZC5C2A77002, 3ZC5C2A85002, 3ZC5C2A85001, 3ZC5C2A74005, 3ZC5C2A72005, 3ZC5C2A73002, 3ZC5C2A73009, 3ZC5C2A73010, 3ZC5C2A74007, 3ZC5C2A76008, 3ZC5C2A76009, 3ZC5C2A76017, 3ZC5C2A77001, 3ZC5C2A77009, 3ZC5C2A77017, 3ZC5C2A78005, 3ZC5C2A78006, 3ZC5C2A79007, 3ZC5C2A7B008, 3ZC5C2A7B011, 3ZC5C2A7B016, 3ZC5C2A7C014, 3ZC5C2A7C017, 3ZC5C2A81005, 3ZC5C2A81006, 3ZC5C2A81014, 3ZC5C2A84009, 3ZC5C2A85011, 3ZC5C2A85013, 3ZC5C2A75008, 3ZC5C2A85009, 3ZC5C2A7B009, 3ZC5C2A76001, 3ZC5C2A7C009, 3ZC5C2A7C010, 3ZC5C2A87009, 3ZC5C2A7C005, 3ZC5C2A7B013, 3ZC5C2A7B014, 3ZC5C2A73003, 3ZC5C2A77005, 3ZC5C2A74006, 3ZC5C2A81007, 3ZC5C2A75003, 3ZC5C2A72002, 3ZC5C2A72003, 3ZC5C2A72004, 3ZC5C2A72006, 3ZC5C2A73011, 3ZC5C2A74002, 3ZC5C2A75004, 3ZC5C2A76007, 3ZC5C2A76014, 3ZC5C2A76015, 3ZC5C2A77007, 3ZC5C2A77008, 3ZC5C2A77015, 3ZC5C2A77016, 3ZC5C2A77019, 3ZC5C2A77022, 3ZC5C2A79004, 3ZC5C2A7B006, 3ZC5C2A7B012, 3ZC5C2A7B018, 3ZC5C2A7C008, 3ZC5C2A81008, 3ZC5C2A82001, 3ZC5C2A84010, 3ZC5C2A84011, 3ZC5C2A85006, 3ZC5C2A85007, 3ZC5C2A86007, 3ZC5C2A87007, 3ZC5C2A88002, 3ZC5C2A88006, 3ZC5C2A87011, 3ZC5C2A87012, 3ZC5C2A86012, 3ZC5C2A77004, 3ZC5C2A7B003, 3ZC5C2A74008, 3ZC5C2A75001, 3ZC5C2A76011, 3ZC5C2A77021, 3ZC5C2A7B004, 3ZC5C2A7B007, 3ZC5C2A7B010, 3ZC5C2A7C015, 3ZC5C2A85012, 3ZC5C2A86005, 3ZC5C2A87002, 3ZC5C2A87003, 3ZC5C2A76024, 3ZC5C2A77018, 3ZC5C2A79006, 3ZC5C2A81003, 3ZC5C2A82002, 3ZC5C2A86008, 3ZC5C2A87013, 3ZC5C2A88001, 3ZC5C2A77020, 3ZC5C2A83001, 3ZC5C2A85004, 3ZC5C2A85008, 3ZC5C2A73005, 3ZC5C2A78004, 3ZC5C2A79005, 3ZC5C2A7B017, 3ZC5C2A7C001, 3ZC5C2A82003, 3ZC5C2A83004, 3ZC5C2A84001, 3ZC5C2A74010, 3ZC5C2A72001, 3ZC5C2A73004, 3ZC5C2A73008, 3ZC5C2A87005, 3ZC5C2A76004, 3ZC5C2A76025, 3ZC5C2A7A003, 3ZC5C2A7C004, 3ZC5C2A7C011, 3ZC5C2A81011, 3ZC5C2A81013, 3ZC5C2A84003, 3ZC5C2A74001, 3ZC5C2A87004, 3ZC5C2A87006, 3ZC5C2A88004, 3ZC5C2A88007, 3ZC5C2A74003, 3ZC5C2A75010, 3ZC5C2A75011, 3ZC5C2A76003, 3ZC5C2A76005, 3ZC5C2A76012, 3ZC5C2A76020, 3ZC5C2A78001, 3ZC5C2A7C003, 3ZC5C2A81010, 3ZC5C2A84005, 3ZC5C2A84012, 3ZC5C2A71002, 3ZC5C2A71004, 3ZC5C2A73007, 3ZC5C2A74009, 3ZC5C2A76023, 3ZC5C2A7C002, 3ZC5C2A73015 Model Numbers: CH-200, UDI codes: N/A, Serial Numbers:, 0162M46304, 0162M47002, 0162M48002, 0162M48801, 0162M48902, 0162M48903, 3ZC5C2A21005, 3ZC5C2A4B008, 3ZC5C2A52005, 3ZC5C2A6B011, 0162M46101, 0162M46104, 0162M47505, 0162M47508, 0162M48702, 0162M49006, 0162M49203, 0162M49501, 0162M50007, 0262M49709, 0262M46103, 3ZC5C2A41006, 0362M46501, 0262M48307, 0362M46710, 0362M48206, 0362M49902, 0462M45303, 0462M48001, 0462M48209, 0462M48302, 0462M48902, 3ZC5C2A16003, 3ZC5C2A21006, 3ZC5C2A23017, 3ZC5C2A23019, 3ZC5C2A28011, 3ZC5C2A31002, 3ZC5C2A32006, 3ZC5C2A34003, 3ZC5C2A35018, 3ZC5C2A38016, 3ZC5C2A39004, 3ZC5C2A39005, 3ZC5C2A3A009, 3ZC5C2A3C010, 3ZC5C2A41001, 3ZC5C2A41002, 3ZC5C2A41003, 3ZC5C2A44006, 3ZC5C2A46006, 3ZC5C2A46007, 3ZC5C2A4C009, 3ZC5C2A53011, 3ZC5C2A53013, 3ZC5C2A5C008, 3ZC5C2A5C009, 3ZC5C2A62003, 3ZC5C2A62004, 3ZC5C2A62008, 3ZC5C2A63011, 3ZC5C2A67010, 3ZC5C2A68002, 3ZC5C2A6C003, 3ZC5C2B16001, 3ZC5C2B31001, 0162M45102, 0162M45124, 0162M45127, 0162M45129, 0162M45130, 0162M45131, 0162M45134, 0162M45135, 0162M45137, 0162M45138, 0162M45144, 0162M45145, 0162M46105, 0162M46106, 0162M46109, 0162M46110, 0162M46210, 0162M46301, 0162M46305, 0162M46605, 0162M46606, 0162M46608, 0162M46708, 0162M47003, 0162M47004, 0162M47306, 0162M47308, 0162M47402, 0162M47404, 0162M47406, 0162M47407, 0162M47509, 0162M47902, 0162M47905, 0162M47909, 0162M48003, 0162M48510, 0162M48802, 0162M48907, 0162M48909, 0162M49005, 0162M49704, 0262M45502, 0262M45503, 0262M45504, 0262M45703, 0262M45704, 0262M45705, 0262M45706, 0262M45710, 0262M45803, 0262M46007, 0262M47207, 0262M47209, 0262M47406, 0262M46104, 0262M47205, 0462M49105, 0262M47505, 0262M47506, 0262M47706, 0262M47707, 0262M47806, 3ZC5C2A45017, 0262M48003, 0362M49305, 0262M48101, 3ZC5C2A2B014, 3ZC5C2A37005, 3ZC5C2A53008, 3ZC5C2A53007, 0262M48107, 0262M48108, 0262M48109, 3ZC5C2A48015, 0262M48208, 0262M48301, 0262M48305, 0262M48306, 0462M48207, 0262M48806, 0462M48301, 0462M48304, 0262M49203, 0262M49204, 0362M46103, 3ZC5C2A45015, 3ZC5C2A57010, 3ZC5C2A31004, 0462M49607, 0262M49210, 0262M49701, 0262M49702, 3ZC5C2A12005, 3ZC5C2A1100A, 3ZC5C2A1A004, 3ZC5C2A0A00A, 0262M49808, 0262M48106, 3ZC5C2A0C005, 0462M50002, 3ZC5C2A16007, 0462M46401, 3ZC5C2A63010, 3ZC5C2A26002, 0362M45403, 0362M45404, 0462M49104, 3ZC5C2A28007, 3ZC5C2A1C009, 3ZC5C2A14007, 3ZC5C2B22001, 0362M50005, 0362M45705, 3ZC5C2A2B015, 0362M46102, 0262M47408, 3ZC5C2A37012, 3ZC5C2A23003, 0362M49905, 0462M47601, 3ZC5C2A47010, 0362M48203, 0362M47109, 0362M48304, 0362M46603, 0362M46708, 0362M47501, 0362M47902, 0362M48204, 0362M46707, 0362M50003, 0362M48305, 0562M45606, 3ZC5C2A33008, 3ZC5C2A1C004, 3ZC5C2A27001, 3ZC5C2A18008, 3ZC5C2A36006, 3ZC5C2A16009, 0462M47409, 3ZC5C2A36004, 3ZC5C2A66011, 3ZC5C2A2C001, 0362M47502, 0362M47503, 0262M48804, 0362M49709, 0462M46101, 0262M49001, 0262M49006, 0362M48302, 3ZC5C2A0A009, 0362M47507, 0362M47903, 3ZC5C2A0A003, 3ZC5C2A62005, 0462M49006, 3ZC5C2A16010, 3ZC5C2A51005, 3ZC5C2A46008, 3ZC5C2B2B003, 0362M45401, 0362M46607, 3ZC5C2A2C010, 0262M48202, 0362M49306, 3ZC5C2A0A007, 0462M49609, 0362M45704, 0462M47802, 0462M47803, 3ZC5C2A38005, 3ZC5C2A2C012, 0362M47104, 0362M49304, 3ZC5C2A47008, 3ZC5C2A57017, 3ZC5C2A64017, 3ZC5C2A6C014, 3ZC5C2A25006, 3ZC5C2A44013, 3ZC5C2A1C008, 3ZC5C2A57015, 3ZC5C2A2C004, 3ZC5C2A27002, 3ZC5C2A19003, 3ZC5C2A49009, 3ZC5C2A12002, 0462M47101, 0362M47908, 0462M47103, 0262M49207, 0262M49209, 0262M49906, 0462M47809, 0462M47704, 0462M46102, 3ZC5C2A21001, 3ZC5C2A61004, 0462M47603, 3ZC5C2A2C002, 0362M46201, 0362M49707, 0362M49906, 0462M45304, 0562M45302, 3ZC5C2A38003, 0462M46605, 0462M47705, 3ZC5C2A46012, 3ZC5C2A4B010, 3ZC5C2A5B009, 0462M47801, 0262M49206, 0462M47901, 0262M49706, 0262M49907, 0362M45703, 0462M48206, 0362M46709, 0262M49007, 0362M45405, 0362M45410, 0362M47201, 3ZC5C2A66012, 3ZC5C2A49005, 0462M48905, 3ZC5C2A5B008, 3ZC5C2A5A001, 0362M46703, 0362M46706, 3ZC5C2A12007, 0462M49210, 0362M47407, 0362M47410, 0362M49407, 0362M49408, 0462M49606, 3ZC5C2A4A002, 0362M47207, 3ZC5C2A13004, 0362M47202, 0462M46606, 3ZC5C2A1A002, 3ZC5C2A1A003, 0262M47504, 0462M50003, 0562M45608, 0562M46301, 3ZC5C2A0A002, 3ZC5C2A0A006, 3ZC5C2A0B003, 3ZC5C2A13006, 3ZC5C2A14001, 3ZC5C2A16006, 3ZC5C2A16008, 3ZC5C2A17005, 3ZC5C2A19004, 3ZC5C2A1C001, 3ZC5C2A1C002, 3ZC5C2A1C006, 3ZC5C2A21013, 3ZC5C2A22002, 3ZC5C2A22016, 3ZC5C2A22017, 3ZC5C2A23001, 3ZC5C2A24002, 3ZC5C2A25004, 3ZC5C2A26003, 3ZC5C2A28006, 3ZC5C2A2B009, 3ZC5C2A2C003, 3ZC5C2A31003, 3ZC5C2A34002, 3ZC5C2A35019, 3ZC5C2A36003, 3ZC5C2A38004, 3ZC5C2A43003, 3ZC5C2A47006, 3ZC5C2A48005, 3ZC5C2A48020, 3ZC5C2A49001, 3ZC5C2A49006, 3ZC5C2A49007, 3ZC5C2A4A008, 3ZC5C2A4B002, 3ZC5C2A4B007, 3ZC5C2A4B009, 3ZC5C2A54007, 3ZC5C2A55004, 3ZC5C2A56003, 3ZC5C2A56008, 3ZC5C2A5B005, 3ZC5C2A5B010, 3ZC5C2A5C010, 3ZC5C2A63009, 3ZC5C2A64008, 3ZC5C2A66006, 3ZC5C2A66013, 3ZC5C2A67006, 3ZC5C2A6A001, 3ZC5C2A6A007, 3ZC5C2A6A008, 3ZC5C2A6B004, 3ZC5C2A6B005, 3ZC5C2B18001, 3ZC5C2B21001, 3ZC5C2B2B001, 3ZC5C2B2C001, 0162M45123, 0162M46401, 0162M47507, 0162M47907, 0162M48807, 0162M48904, 0162M48906, 0162M50005, 0262M47208, 0262M47503, 0262M48205, 0262M48304, 0362M46110, 0462M47605, 0462M47703, 0462M48208, 0362M47805, 3ZC5C2A24007, 3ZC5C2A23016, 3ZC5C2A18001, 3ZC5C2A23013, 3ZC5C2A35020, 3ZC5C2A5C011, 0562M45603, 0562M45504, 0462M45907, 0462M46402, 3ZC5C2A16005, 3ZC5C2A14003, 3ZC5C2A24001, 3ZC5C2A4C016, 3ZC5C2A5C001, 3ZC5C2A48011, 3ZC5C2A1A005, 3ZC5C2A64020, 3ZC5C2A65007, 3ZC5C2A65008, 3ZC5C2A65009, 3ZC5C2A65011, 3ZC5C2A44007, 0362M47801, 0462M49408, 3ZC5C2A55006, 3ZC5C2A55007, 3ZC5C2A55009, 3ZC5C2A21014, 0462M48003, 0362M47409, 0462M50001, 3ZC5C2A2B016, 3ZC5C2A13003, 3ZC5C2A2B017, 3ZC5C2A2C005, 0462M48303, 3ZC5C2A17006, 3ZC5C2A23009, 3ZC5C2A2C009, 3ZC5C2A21008, 3ZC5C2A21009, 3ZC5C2A33007, 3ZC5C2A36002, 0262M49909, 0362M47408, 0362M47406, 0462M49610, 3ZC5C2A3B001, 3ZC5C2A18011, 3ZC5C2A19001, 3ZC5C2A5A015, 0262M49904, 0562M45306, 3ZC5C2A11006, 3ZC5C2A28009, 0362M49801, 0562M45607, 3ZC5C2A6C012, 3ZC5C2A6C013, 3ZC5C2A22004, 0362M48301, 3ZC5C2A18009, 0462M47604, 3ZC5C2A54008, 3ZC5C2A16001, 3ZC5C2A16002, 3ZC5C2A64005, 0262M49910, 3ZC5C2A5C002, 3ZC5C2A23015, 3ZC5C2A33005, 3ZC5C2A37009, 3ZC5C2A37011, 3ZC5C2A39002, 3ZC5C2A43005, 3ZC5C2A43006, 3ZC5C2A45016, 3ZC5C2A49014, 3ZC5C2A4A003, 3ZC5C2A4C008, 3ZC5C2A55010, 3ZC5C2A5A006, 3ZC5C2A61008, 3ZC5C2A64006, 3ZC5C2B27001, 3ZC5C2B37001, 0162M45119, 0162M45139, 0162M45140, 0162M45141, 0162M46206, 0162M46302, 0162M47005, 0162M47309, 0162M47401, 0162M47601, 0162M48001, 0162M48910, 0162M50003, 0162M50004, 0262M45707, 0262M46009, 0262M46010, 0262M46401, 0262M46403, 0262M46407, 0262M46408, 0262M47204, 0262M47509, 0262M47510, 0262M47801, 0262M47910, 0262M48009, 0262M48904, 0262M48905, 0262M49010, 0262M49208, 0362M46604, 0362M47405, 0362M47803, 0362M49303, 0362M49405, 0362M49508, 0362M49904, 0462M46103, 0462M47702, 0462M48005, 0462M49404, 3ZC5C2A14009, 3ZC5C2A18010, 3ZC5C2A19005, 3ZC5C2A19006, 3ZC5C2A38009, 3ZC5C2A63006, 0362M46305, 3ZC5C2A61005, 3ZC5C2A39003, 3ZC5C2A23021, 3ZC5C2A56006, 0362M48207, 3ZC5C2A32005, 3ZC5C2A22006, 0462M46608, 0462M47810, 0462M46607, 3ZC5C2B28001, 3ZC5C2A52006, 3ZC5C2A52007, 3ZC5C2A22014, 3ZC5C2A55003, 3ZC5C2A49012, 0462M48004, 0362M46901, 0462M48602, 0462M48603, 0462M48604, 0462M48605, 3ZC5C2A23002, 3ZC5C2A23005, 3ZC5C2A6C016, 3ZC5C2A3A010, 3ZC5C2A64013, 3ZC5C2A62013, 3ZC5C2A0C004, 3ZC5C2B12001, 3ZC5C2A3B005, 3ZC5C2A65003, 3ZC5C2A23020, 3ZC5C2A32003, 3ZC5C2A62009, 3ZC5C2A64012, 3ZC5C2A54009, 0262M49903, 3ZC5C2A68004, 3ZC5C2A28005, 3ZC5C2A66010, 3ZC5C2A49008, 3ZC5C2A3C002, 3ZC5C2A67002, 3ZC5C2A11003, 3ZC5C2A3B002, 3ZC5C2A24003, 0362M46108, 3ZC5C2A14002, 3ZC5C2A36010, 3ZC5C2A2C015, 3ZC5C2A1C014, 3ZC5C2A59001, 3ZC5C2A6A005, 0362M46304, 3ZC5C2A3A002, 0462M49101, 0362M48408, 0362M49404, 3ZC5C2A66002, 0562M45101, 0362M48701, 3ZC5C2A43007, 0462M47102, 0562M45304, 0262M49201, 0362M45701, 0362M45702, 0362M47108, 0462M48305, 0262M48207, 0362M47802, 3ZC5C2A36007, 3ZC5C2A3B004, 3ZC5C2A3C004, 3ZC5C2A49002, 3ZC5C2A49010, 3ZC5C2A57003, 3ZC5C2A59004, 3ZC5C2A59010, 3ZC5C2A5A009, 3ZC5C2A5C004, 3ZC5C2A63003, 3ZC5C2A63005, 3ZC5C2A64007, 3ZC5C2A64011, 3ZC5C2A66005, 3ZC5C2A66007, 3ZC5C2A66008, 3ZC5C2A6A002, 3ZC5C2A6A004, 3ZC5C2A6A012, 3ZC5C2A6B002, 3ZC5C2A6B008, 3ZC5C2A6C001, 3ZC5C2A6C005, 3ZC5C2A6C011, 3ZC5C2A6C015, 0162M45118, 0162M45120, 0162M46402, 0162M46603, 0162M47510, 0162M48808, 0162M49001, 0162M49003, 0162M49105, 0162M49204, 0162M49210, 0162M50001, 0166M50001, 0262M45905, 0262M46402, 0262M47202, 0262M47203, 0262M47810, 0262M47906, 0262M48002, 0262M48005, 0262M48201, 0262M48805, 0262M49005, 0262M49009, 0262M49205, 0362M46107, 0362M46705, 0362M47906, 0362M49509, 0462M45906, 0462M49110, 0462M49405, 0462M49902, 0562M45605, 3ZC5C2A0A004, 3ZC5C2A0C001, 3ZC5C2A19007, 3ZC5C2A1B001, 3ZC5C2A1B002, 3ZC5C2A1B007, 3ZC5C2A1B012, 3ZC5C2A1B013, 3ZC5C2A22001, 3ZC5C2A23008, 3ZC5C2A23014, 3ZC5C2A24006, 3ZC5C2A24009, 3ZC5C2A27003, 3ZC5C2A2B006, 3ZC5C2A31005, 3ZC5C2A31009, 3ZC5C2A32004, 3ZC5C2A36001, 0562M45604, 3ZC5C2A51004, 3ZC5C2A37006, 0562M45807, 0462M45905, 0462M47808, 3ZC5C2A69007, 3ZC5C2A21017, 3ZC5C2A64009, 3ZC5C2A37019, 3ZC5C2A38001, 3ZC5C2A5A007, 3ZC5C2A63007, 3ZC5C2A19002, 3ZC5C2A21016, 3ZC5C2A1A001, 3ZC5C2A38010, 3ZC5C2A18003, 3ZC5C2A18002, 3ZC5C2A38014, 3ZC5C2A38015, 3ZC5C2A26004, 3ZC5C2A37018, 3ZC5C2A44005, 3ZC5C2A3A005, 3ZC5C2A3A006, 3ZC5C2A6A006, 3ZC5C2A3B006, 3ZC5C2A3B007, 3ZC5C2A2B007, 3ZC5C2A36005, 3ZC5C2A22008, 3ZC5C2A37003, 3ZC5C2A37004, 3ZC5C2A22007, 3ZC5C2A18004, 0362M50002, 3ZC5C2A38008, 0462M47407, 3ZC5C2A14008, 3ZC5C2A46010, 3ZC5C2A62011, 3ZC5C2A43004, 3ZC5C2A65010, 3ZC5C2A61010, 0462M47902, 3ZC5C2A33004, 3ZC5C2A52008, 3ZC5C2A31007, 3ZC5C2A45009, 3ZC5C2A2C014, 3ZC5C2A31006, 3ZC5C2A5A003, 3ZC5C2A44014, 3ZC5C2A21003, 3ZC5C2A67003, 3ZC5C2A28010, 3ZC5C2A63008, 3ZC5C2A38002, 3ZC5C2A35022, 3ZC5C2A45019, 3ZC5C2A46004, 3ZC5C2A46011, 3ZC5C2A5A008, 3ZC5C2A5B007, 3ZC5C2A48003, 3ZC5C2A56002, 0462M47410, 3ZC5C2A47003, 0562M45602, 3ZC5C2A23004, 0262M48105, 0462M49904, 0262M48104, 3ZC5C2A12003, 0462M49905, 0462M49901, 3ZC5C2A39001, 3ZC5C2A48006, 3ZC5C2A67008, 0462M47610, 3ZC5C2A5A002, 0362M47709, 3ZC5C2A54011, 3ZC5C2A56001, 3ZC5C2A48017, 3ZC5C2A1C007, 3ZC5C2A5A012, 3ZC5C2A69013, 3ZC5C2A52001, 3ZC5C2A54001, 3ZC5C2A69010, 3ZC5C2A0B006, 3ZC5C2A52003, 3ZC5C2A52004, 3ZC5C2A5C005, 3ZC5C2A5C006, 3ZC5C2A52002, 0562M45301, 3ZC5C2A33003, 3ZC5C2A4A007, 3ZC5C2A52010, 3ZC5C2A4A010, 3ZC5C2A57005, 0262M48010, 0462M47408, 3ZC5C2A37013, 3ZC5C2A22005, 3ZC5C2A23018, 3ZC5C2A55005, 3ZC5C2A37020, 3ZC5C2A3A003, 3ZC5C2A14005, 3ZC5C2B11002, 3ZC5C2A23011, 0262M49707, 3ZC5C2A37014, 3ZC5C2A56010, 3ZC5C2A2B002, 3ZC5C2A56004, 3ZC5C2A56005, 3ZC5C2A53009, 3ZC5C2A53010, 3ZC5C2A53006, 3ZC5C2A51003, 3ZC5C2A57013, 3ZC5C2A47012, 3ZC5C2A5C003, 3ZC5C2A5A011, 3ZC5C2A49011, 0462M49102, 3ZC5C2A57001, 3ZC5C2A57004, 3ZC5C2A5A010, 3ZC5C2A5B001, 3ZC5C2A62010, 3ZC5C2A62012, 3ZC5C2A63013, 3ZC5C2A63015, 3ZC5C2A66009, 3ZC5C2A6B010, 3ZC5C2A6C004, 3ZC5C2B27002, 3ZC5C2B2B002, 3ZC5C2B32001, 0262M45506, 0262M45508, 0262M45903, 0262M47805, 0262M47807, 0262M48204, 0262M49710, 0362M49704, 0562M45808, 3ZC5C2A11008, 3ZC5C2A1C011, 3ZC5C2A48016, 0262M47407, 0462M49004, 0362M45402, 3ZC5C2A2C016, 0262M48807, 0362M47907, 3ZC5C2A22012, 0362M49307, 3ZC5C2A3C008, 3ZC5C2A22011, 3ZC5C2A36011, 3ZC5C2A28008, 3ZC5C2A48001, 3ZC5C2A47017, 3ZC5C2A2C006, 3ZC5C2A37008, 3ZC5C2A51002, 3ZC5C2A0C003, 3ZC5C2A51001, 3ZC5C2A26001, 3ZC5C2A57007, 0262M48709, 3ZC5C2A44015, 3ZC5C2A63001, 3ZC5C2A64016, 3ZC5C2A6C002, 0162M47405, 0162M48901, 0162M48908, 0262M46901, 0262M46903, 0262M47904, 3ZC5C2A11004, 3ZC5C2A11005, 3ZC5C2A55002, 3ZC5C2A65002, 3ZC5C2A4B003, 3ZC5C2A21007, 3ZC5C2A54006, 3ZC5C2A38006, 3ZC5C2A54003, 3ZC5C2B34001, 3ZC5C2B26001, 3ZC5C2A25003, 3ZC5C2A27004, 0462M47804, 3ZC5C2A45014, 3ZC5C2A0A008, 0462M49310, 3ZC5C2A4A009, 3ZC5C2A56007, 0262M49004, 0362M49705, 3ZC5C2A17003, 3ZC5C2A17004, 3ZC5C2A2B011, 0462M45301, 0462M45302, 0362M46101, 3ZC5C2A47007, 3ZC5C2A25002, 3ZC5C2B1B001, 3ZC5C2A22010, 0462M45904, 3ZC5C2B24001, 3ZC5C2A65001, 3ZC5C2A66004, 3ZC5C2B27003, 0462M49008, 3ZC5C2A53014, 3ZC5C2A67001, 0262M46406, 0362M47110, 0362M45503, 0262M45801, 0262M46701, 0262M47809, 0462M45902, 0562M45102, 3ZC5C2A2C013, 3ZC5C2A3C001, 0362M47904, 0362M47710, 3ZC5C2A54002, 3ZC5C2A31001, 3ZC5C2B1A001, 3ZC5C2B32002, 3ZC5C2A46009, 0262M47201, 3ZC5C2A6B006, 3ZC5C2A67004, 3ZC5C2A6B007, 3ZC5C2B1C001, 0162M46303, 0162M46604, 0262M47405, 0462M45308, 3ZC5C2A18006, 3ZC5C2A24012, 3ZC5C2A62006, 3ZC5C2A47015, 3ZC5C2A49013, 3ZC5C2A5B004, 3ZC5C2A13005, 3ZC5C2A3C007, 3ZC5C2A24011, 0262M45507, 0262M46902, 3ZC5C2A5B006, 3ZC5C2B16002, 0162M46607, 0162M46705, 0162M47001, 0162M47301, 0162M47310, 0162M50008, 0262M47206, 0362M46605, 3ZC5C2A41004, 3ZC5C2A43009, 3ZC5C2A66001, 3ZC5C2A41005, 0362M45706, 0362M49308, 3ZC5C2A31008, 3ZC5C2A1C013, 0462M49010, 3ZC5C2A43008, 3ZC5C2A48012, 3ZC5C2A64015, 3ZC5C2A48014, 3ZC5C2A1B004, 3ZC5C2A12006, 3ZC5C2A0A00B, 3ZC5C2A71001, 0162M45122, 0162M46207, 0162M46710, 0162M47305, 0262M48308, 0262M48707, 0462M46309, 0462M49706, 0562M45505, 3ZC5C2A15001, 3ZC5C2A32002, 3ZC5C2A33001, 3ZC5C2A34009, 3ZC5C2A55008, 3ZC5C2A69008, 0162M45136, 0162M46509, 0162M46706, 0162M47307, 0162M47904, 0162M48810, 0162M49007, 0162M49909, 0262M42808, 0262M45708, 0262M45802, 0262M46006, 0262M46008, 0262M47104, 0262M47802, 0262M47808, 0262M48006, 0262M49202, 0362M46902, 0362M50001, 0462M46403, 0462M49307, 3ZC5C2A18007, 3ZC5C2A2C011, 3ZC5C2A37001, 3ZC5C2A46005, 3ZC5C2A54010, 3ZC5C2A57002, 3ZC5C2A59003, 3ZC5C2A6A003, 3ZC5C2A6C007, 3ZC5C2A1C012, 3ZC5C2B11001, 3ZC5C2A38013, 0262M49003, 3ZC5C2A54005, 3ZC5C2A63014, 3ZC5C2A6C010, 3ZC5C2A6C008, 3ZC5C2A35023, 0462M49608, 3ZC5C2A69011, 3ZC5C2A48023, 3ZC5C2A48013, 3ZC5C2A4B001, 3ZC5C2A4A004, 3ZC5C2A4A001, 0562M45305, 3ZC5C2A13002, 3ZC5C2A4C011, 3ZC5C2A2B005, 3ZC5C2A2B013, 3ZC5C2A16004, 3ZC5C2A3C009, 3ZC5C2A61002, 3ZC5C2A1C003, 0362M49507, 0462M45306, 3ZC5C2A1B005, 0462M49308, 0562M45105, 0562M45601, 3ZC5C2A22013, 0562M45609, 0562M45610, 3ZC5C2A62007, 3ZC5C2A37007, 3ZC5C2A33002, 3ZC5C2A47011, 3ZC5C2A45010, 3ZC5C2A41008, 3ZC5C2A41013, 3ZC5C2A3C005, 3ZC5C2A44010, 3ZC5C2A44001, 3ZC5C2A44009, 3ZC5C2A1B011, 3ZC5C2A24005, 3ZC5C2A6B009, 3ZC5C2A35016, 0162M45101, 0162M45121, 0162M45125, 0162M45126, 0162M45132, 0162M45133, 0162M45142, 0162M45143, 0162M46102, 0162M46103, 0162M46707, 0162M46709, 0162M47602, 0162M47607, 0162M47908, 0162M47910, 0162M49104, 0162M49502, 0162M49503, 0162M49602, 0162M50006, 0262M42802, 0262M45505, 0262M45709, 0262M45904, 0262M45906, 0262M46703, 0262M47409, 0262M47705, 0262M47908, 0262M48102, 0262M48710, 0262M48808, 0262M49809, 0362M46004, 0362M47103, 0362M47701, 0362M48303, 0362M49505, 0362M49701, 0462M45307, 0562M45104, 3ZC5C2A12009, 3ZC5C2A14010, 3ZC5C2A21004, 3ZC5C2A2B008, 3ZC5C2A39006, 3ZC5C2A3C006, 3ZC5C2A44011, 3ZC5C2A44012, 3ZC5C2A47009, 3ZC5C2A47013, 3ZC5C2A53015, 3ZC5C2A5A014, 3ZC5C2A64010, 3ZC5C2A64014, 3ZC5C2A64019, 3ZC5C2A67005, 3ZC5C2A67007, 3ZC5C2A69006, 3ZC5C2A69012, 3ZC5C2A6A009, 3ZC5C2A6B001, 3ZC5C2A6C009, 0262M49908, 0362M49901, 0462M49406, 3ZC5C2A12008, 3ZC5C2A1200A, 3ZC5C2A1200B, 3ZC5C2A14006, 0462M45903, 3ZC5C2B19001, 3ZC5C2A23007, 0362M49702, 3ZC5C2A5B011, 0362M49703, 3ZC5C2A41007, 3ZC5C2A48008, 3ZC5C2A6C006, 0462M49009, 3ZC5C2A5A004, 0362M46109, 0362M48009, 3ZC5C2A45013, 3ZC5C2A61006, 3ZC5C2A61007, 3ZC5C2A63017, 0262M47907, 3ZC5C2A47005, 3ZC5C2A25001, 3ZC5C2A48019, 3ZC5C2A4A006, 3ZC5C2A57006, 0362M50004, 0462M47105, 0462M47107, 0362M47508, 3ZC5C2A37015, 3ZC5C2A37016, 3ZC5C2A31011, 0462M47104, 3ZC5C2A1C005, 3ZC5C2A17002, 0462M46206, 3ZC5C2A0B002, 0462M47805, 3ZC5C2A0B005, 3ZC5C2A11007, 3ZC5C2A11009, 0262M49708, 3ZC5C2A6C017, 0362M47105, 3ZC5C2A58001, 3ZC5C2A12004, 0462M49903, 3ZC5C2A1B010, 3ZC5C2A21011, 0462M47706, 0462M47701, 0362M46602, 0362M46606, 0362M49504, 0462M47106, 0362M46106, 0362M49708, 0462M47108, 0262M47909, 3ZC5C2A5C007, 3ZC5C2A56009, 0262M48110, 0462M45305, 0462M49309, 3ZC5C2A59002, 3ZC5C2A41010, 3ZC5C2A41012, 3ZC5C2A41009, 3ZC5C2A41011, 3ZC5C2A44003, 3ZC5C2A44002, 3ZC5C2A31010, 3ZC5C2A63016, 3ZC5C2A63018, 3ZC5C2A64004, 3ZC5C2A52009, 3ZC5C2A22015, 3ZC5C2A33006, 3ZC5C2A0B004, 3ZC5C2A58002, 0362M49506, 3ZC5C2A38007, 3ZC5C2A3A001, 0362M46601, 0362M46704, 3ZC5C2A1B009, 0562M45303, 3ZC5C2A22003, 3ZC5C2A48018, 3ZC5C2A1B014, 3ZC5C2A1B008, 0362M47804, 0362M47905, 3ZC5C2A4A005, 3ZC5C2A48010, 3ZC5C2A4B004, 3ZC5C2A2B010, 3ZC5C2A31012, 3ZC5C2A32001, 3ZC5C2A34001, 3ZC5C2A45011, 3ZC5C2A47002, 3ZC5C2A54004, 3ZC5C2A23010, 3ZC5C2A48004, 3ZC5C2A65006, 3ZC5C2A35017, 3ZC5C2A69009, 0262M48303, 3ZC5C2A47004, 3ZC5C2A24010, 0262M49002, 0462M47707, 0462M47708, 3ZC5C2A63002, 3ZC5C2A48022, 3ZC5C2A2B003, 0462M46302, 3ZC5C2A47016, 0462M49007, 0362M49301, 3ZC5C2A48007, 3ZC5C2A21002, 3ZC5C2A38012, 0262M46702, 0262M48103, 0262M48302 Model Numbers: CH-200M, UDI codes: 04540217050826, Serial Numbers:, 41E733E74005, 41E733E72003, 41E733E85002, 41E733E86001, 41E733E86002, 41E733E86003, 41E733E86004, 41E733E88001, 41E733E72001, 41E733E72002, 41E733E72005, 41E733E74001, 41E733E74003, 41E733E74004, 41E733E85001, 41E733E88002, 41E733E88003, 41E733E72004, 41E733E86005, 41E733E74002, 41E733E88004 Model Numbers: CH-200M, UDI codes: N/A, Serial Numbers:, 0262M42603, 0362M44102, 0462M43402, 0462M43403, 0562M43907, 0362M44701, 41E733E67002, 41E733E64004, 41E733E65001, 41E733E63005, 3ZC5C1661001, 41E733E64001, 41E733E6B004, 0462M40101, 3ZC5C160C005, 0662M41305, 0562M42808, 3ZC5C1653001, 3ZC5C1652002, 0662M40807, 0662M40806, 0662M40401, 0362M44502, 0462M43404, 3ZC5C1632004, 0262M40510, 3ZC5C1617001, 0462M45001, 0462M45003, 0562M44901, 3ZC5C1645001, 3ZC5C1645002, 3ZC5C1612005, 3ZC5C162C001, 3ZC5C162A001, 0662M41303, 0562M42805, 3ZC5C1622001, 0562M40701, 3ZC5C160C007, 3ZC5C1635002, 0362M44702, 0362M44703, 3ZC5C162B002, 0262M44906, 0162M43610, 0162M43613, 0162M43618, 0162M43619, 0162M43620, 0162M43916, 0162M44409, 0162M44509, 0262M40509, 0262M40610, 0262M41609, 0262M43904, 0262M44301, 0262M44302, 0262M44303, 0262M44408, 0362M40801, 0362M40802, 0362M40803, 0362M41407, 0362M41409, 0362M42404, 0462M40103, 0462M41202, 0562M42806, 0662M40808, 3ZC5C160C006, 3ZC5C1612002, 3ZC5C1613002, 3ZC5C1615001, 3ZC5C161B003, 3ZC5C1623002, 3ZC5C1631002, 3ZC5C1643002, 3ZC5C1643003, 3ZC5C1643005, 41E733E63001, 41E733E67001, 41E733E67005, 41E733E6B001, 41E733E6B009, 0462M40102, 3ZC5C160C004, 41E733E67004, 3ZC5C1645004, 3ZC5C162A002, 0462M41201, 3ZC5C1616004, 3ZC5C162B001, 3ZC5C1636001, 3ZC5C160C003, 0662M40809, 3ZC5C1616001, 3ZC5C1654001, 0462M41602, 3ZC5C1629005, 3ZC5C161A001, 0462M40202, 3ZC5C162A004, 0362M40510, 0462M44204, 3ZC5C1646002, 0162M43015, 0162M43016, 0162M43218, 0162M43614, 0262M40507, 0262M40710, 0262M42604, 0362M42405, 0562M43906, 3ZC5C164C002, 0362M43603, 3ZC5C1612004, 41E733E69005, 3ZC5C161A002, 3ZC5C1635001, 3ZC5C1629003, 3ZC5C1623003, 3ZC5C1629001, 3ZC5C1616003, 0462M42002, 3ZC5C1632002, 0162M43017, 0162M43612, 0262M40508, 0262M40607, 0262M43602, 0462M41701, 0462M43202, 0462M45002, 0562M40702, 0562M43110, 3ZC5C1638001, 3ZC5C164A002, 3ZC5C1655002, 41E733E6B003, 3ZC5C164A001, 41E733E64003, 3ZC5C161B001, 3ZC5C1625001, 3ZC5C1648001, 0362M44501, 3ZC5C1659001, 41E733E63003, 3ZC5C1652001, 3ZC5C1641001, 0362M44704, 3ZC5C1631001, 3ZC5C160C001, 3ZC5C1612003, 0162M42920, 0162M43611, 0162M43616, 0262M40506, 0262M42801, 0262M42901, 0462M41204, 0462M43001, 0462M44202, 3ZC5C1629004, 3ZC5C1632003, 3ZC5C1633001, 41E733E6A001, 41E733E6B002, 41E733E69002, 0562M42608, 41E733E64002, 3ZC5C164C001, 0262M44403, 0362M43601, 3ZC5C1637001, 3ZC5C1637002, 41E733E63002, 3ZC5C162C002, 3ZC5C162A003, 0462M41203, 0362M44104, 0462M41601, 0262M44205, 0462M40201, 0462M44203, 0262M44409, 0462M42001, 0362M41408, 0662M41304, 0162M43615, 0262M43202, 0262M43510, 0262M44407, 3ZC5C1632001, 3ZC5C1628002, 0462M43002, 0162M43617, 0462M42402, 3ZC5C1616002, 0562M42804, 3ZC5C1612006, 3ZC5C1641002, 3ZC5C1623004, 3ZC5C1637003, 3ZC5C161B002, 3ZC5C1612008, 0562M42807, 0362M43002, 0262M42709, 0262M44304, 0262M44204, 0362M44905, 3ZC5C1612007, 3ZC5C1655001, 3ZC5C1661002, 41E733E69004, 0462M43401, 0462M43406, 0462M42104, 0462M43405, 0462M44002, 0462M45004, 0462M43201, 0162M43518, 0162M44510, 0262M43905, 3ZC5C1633002, 0362M41101, 0362M42303, 0362M42403, 0362M43604, 0362M44103, 3ZC5C1636002, 3ZC5C1629002, 3ZC5C1643004, 3ZC5C1613001, 0462M43803, 0462M44205, 0562M41105, 0662M40601, 3ZC5C1612001, 3ZC5C1623001, 0562M42803, 41E733E63004, 41E733E69001, 3ZC5C163A001, 3ZC5C1645003, 3ZC5C1648002, 0462M42105, 41E733E67003, 0462M41205, 3ZC5C1717001, 3ZC5C1628001, 41E733E6B006, 0262M44203, 0262M44807, 41E733E6B005, 41E733E6B007, 41E733E6B008.
Why was this product recalled?
Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating shaft.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION take?
On 10/16/2025, the firm emailed an "Urgent: Voluntary Medical Device Recall Notice" to customers informing them that, the firm has identified potential problems with the Ceiling Tube Support that consists of an X-ray Tube holder with has a rotation mechanism. There is the potential for rotation shaft cracking due to periodic movements. If the cracks grow, the retaining shaft may eventually break. The mechanical stopper is equipped to prevent X-ray tube assembly from coming off even if the retaining shaft breaks. It may not work effectively only if the shaft breaks while being operated toward front while lifting. X-ray tube assembly weighs 40 kg and if it were to fall off and contact the patient or the operator, it could cause critical injuries. Customer are instructed to: - Perform the daily inspection described in the CH-200/CH-200M operation manual to check for wobbles or shakes, misalignment of the light irradiation position, etc. If any abnormality is found, discontinue use and contact our service provider. When rotating the X-ray tube unit, do not hit the stopper at the end. If the stopper is hit hard, in the unlikely event that a crack has already occurred, a large impact may be applied, and the growth of the crack may be accelerated. In the case of electric rotation, it is controlled by software and does not hit the stopper. - Do not rotate the X-ray tube over or near the patient. For questions - Shimadzu Medical Systems National Technical Support: Toll Free: (844)487-2767 Available 8:00 to 5:00 PM, Monday - Friday
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0499-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology