VerifyNow PRUTest Platelet Reactivity Test REF 85225 UDI-DI code: 10711234150078 The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. Recall
VerifyNow PRUTest Platelet Reactivity Test REF 85225 UDI-DI code: 10711234150078 The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade.
FDA recall Class III
Ongoing
FDA lists this Class III recall for VerifyNow PRUTest Platelet Reactivity Test REF 85225 UDI-DI code: 10711234150078 The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0500-2026
- Recall classification
- Class III
- Enforcement status
- Ongoing
- Initiation date
- Oct 21, 2025
- FDA report date
- Nov 19, 2025
- Quantity in commerce
- 125 boxes (3,125 tests)
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
VerifyNow PRUTest Platelet Reactivity Test REF 85225 UDI-DI code: 10711234150078 The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade.
Affected model & UDI checker
Search the exact identifiers published for Z-0500-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 85225 | VerifyNow PRUTest Platelet Reactivity Test REF 85225 UDI-DI code: 10711234150078 The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. | Check source notice | Not specified |
| Not specified | VerifyNow PRUTest Platelet Reactivity Test REF 85225 UDI-DI code: 10711234150078 The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. | Check source notice | 10711234150078 |
FDA code information
Catalog Number: 85225 UDI-DI code: 10711234150078 Lot Number: WP1559-01
Why was this product recalled?
Due to an device without a premarket clearance being incorrectly package and distributed.
What actions did Accriva Diagnostics, Inc. take?
On 10/21/2025, the firm sent via email an "Urgent Medical Device Recall" Letter to customers informing them that Accriva Diagnostics, Inc. has identified that a small number of VerifyNow PRUTest Platelet Reactivity Test (85225) cartridge pouches from lot WP1559-01 may contain VerifyNow P2Y12 Test cartridges. VerifyNow P2Y12 Test cartridges are not cleared by the US-FDA for use in the United States. Customers are instructed to: 1. Please forward this communication to those within your organization who need to become aware of this matter and verify they understand the procedure and can complete the enclosed forms. 2. In addition, forward the enclosed customer notification to all other organizations where affected devices may have been transferred to. 3. Please check your inventory to determine if you have any boxes of VerifyNow PRU Platelet Reactivity Test (85225), lot number WP1559-01. If you have this product in stock, do NOT use. 4. Please sort through your inventory to identify if your inventory is affected. 5. Please destroy all impacted boxes and document on the enclosed form. 6. Please read the instructions on the Tracking Form and complete and return the form to Accriva Diagnostics, Inc. 7. If you have questions regarding replacement product, please contact Customer Service at 1-800-955-9525 or e-mail [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0500-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026