MC3 VitalFlow Console, REF 58100; Blood pump of ecmo Recall
MC3 VitalFlow Console, REF 58100; Blood pump of ecmo
FDA recall Class II
Ongoing
FDA lists this Class II recall for MC3 VitalFlow Console, REF 58100; Blood pump of ecmo. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0509-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 3, 2025
- FDA report date
- Nov 26, 2025
- Quantity in commerce
- 167 units
- Distribution
- See key details
- FDA cause
- Software in the Use Environment
What is being recalled?
MC3 VitalFlow Console, REF 58100; Blood pump of ecmo
Affected model & UDI checker
Search the exact identifiers published for Z-0509-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 58100 | MC3 VitalFlow Console, REF 58100; Blood pump of ecmo | Check source notice | Not specified |
| Not specified | MC3 VitalFlow Console, REF 58100; Blood pump of ecmo | Check source notice | 10850007732247 |
FDA code information
UDI/DI 10850007732247, Serial Numbers: 000023, 000047, 000073, 000100, 000126, 000150, 000173, 000197, 000223, 000024, 000048, 000075, 000101, 000127, 000151, 000174, 000202, 000224, 000025, 000051, 000076, 000102, 000128, 000152, 000177, 000203, 000225, 000027, 000052, 000078, 000103, 000129, 000153, 000179, 000204, 000228, 000028, 000054, 000079, 000104, 000130, 000154, 000180, 000205, 000229, 000029, 000055, 000080, 000106, 000131, 000155, 000181, 000206, 000277, 000030, 000056, 000081, 000108, 000132, 000156, 000182, 000207, 000284, 000031, 000058, 000082, 000109, 000133, 000157, 000183, 000208, 000032, 000059, 000083, 000110, 000134, 000159, 000184, 000209, 000033, 000060, 000084, 000112, 000135, 000161, 000185, 000210, 000034, 000061, 000085, 000115, 000136, 000162, 000186, 000211, 000035, 000063, 000087, 000116, 000140, 000163, 000187, 000212, 000037, 000065, 000088, 000117, 000141, 000165, 000188, 000213, 000038, 000066, 000089, 000118, 000142, 000166, 000189, 000214, 000039, 000067, 000090, 000119, 000143, 000167, 000190, 000215, 000040, 000068, 000092, 000120, 000145, 000168, 000191, 000216, 000043, 000069, 000095, 000121, 000146, 000169, 000192, 000217, 000044, 000070, 000096, 000122, 000147, 000170, 000193, 000218, 000045, 000071, 000097, 000124, 000148, 000171, 000194, 000219, 000046, 000072, 000099, 000125, 000149, 000172, 000195, 000221.
Why was this product recalled?
As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.
What actions did MEDTRONIC PERFUSION SYSTEMS take?
Medtronic Cardiac Surgery issued an Urgent: Medical Device Correction notice to its consignees on 10/03/2025. The notice explained the issue, potential harms, and requested the consignee contact Medtronic Sales Representative to schedule the service event to their affected VitalFlow" Console. Please reference Technical Service Update (TSU) "MOD0209" when scheduling. The notice also indicated that the consignees use unaffected VitalFlow Console devices, if available, until the correction has been completed. For questions regarding the communication, please contact your Medtronic Sales Representative.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0509-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology