FDA RECALL RECORD · Z-0511-2025

MEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622A Recall

MEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622A

FDA recall Class II

Ongoing

FDA lists this Class II recall for MEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622A. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0511-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 16, 2024
FDA report date
Dec 4, 2024
Quantity in commerce
3396 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

MEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622A

Affected model & UDI checker

Search the exact identifiers published for Z-0511-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
DYNJRA9043SMEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622ACheck source noticeNot specified
PAIN1622AMEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622ACheck source noticeNot specified
Not specifiedMEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622ACheck source notice10193489888287
Not specifiedMEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622ACheck source notice40193489888288
Not specifiedMEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622ACheck source notice10193489891690
Not specifiedMEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622ACheck source notice40193489891691
FDA code information

Codes: 1) DYNJRA9043S, UDI/DI 10193489888287(each) 40193489888288(case), Lot Numbers: 23CBE390, 23CBO216, 23EBM309, 23EBP979, 23EBS281, 23GBV099, 23HBG915, 23IBG610, 23IBG611, 23JBQ712, 23KBC341; 2) PAIN1622A, UDI/DI 10193489891690 (each) 40193489891691 (case), Lot Numbers: 21ELA821, 21GLA615, 21KLA308, 22ALA608, 22CLA681, 22ELB088, 22FLB075, 22GLA543, 22KLA019, 23BLA333, 23DLA466, 23HLA896

View the original FDA recall notice

Why was this product recalled?

The connectors may not allow catheters to be inserted as intended if the connector is not in the fully open position.

What actions did MEDLINE INDUSTRIES, LP - Northfield take?

An IMMEDIATE ACTION REQUIRED RECALL notification letter dated 10/16/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submitted response form, your account will receive over-labels to place unaffected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.comRecall Reference #: R-24-216-FGRecall Code: 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.