MEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597 Recall
MEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597
FDA recall Class II
Ongoing
FDA lists this Class II recall for MEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0518-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 30, 2025
- FDA report date
- Nov 26, 2025
- Quantity in commerce
- 12 kits
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
MEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597
Affected model & UDI checker
Search the exact identifiers published for Z-0518-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| DYNJ81597 | MEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597 | Check source notice | Not specified |
| Not specified | MEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597 | Check source notice | 10195327124007 |
| Not specified | MEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597 | Check source notice | 40195327124008 |
FDA code information
UDI/DI 10195327124007 (each), 40195327124008 (case), Lot Numbers: 23JBR560
Why was this product recalled?
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Medline Industries, LP take?
Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/30/2025 and 10/6/2025 via email and mail. The notice explained the issue with the component, potential risk, and requested the following actions: 1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information found in the notice to complete the response form. Please list the quantity of affected product in inventory on the form. Even if no affected product is in inventory, please complete and submit the form. 3. Upon receipt of the submitted response form, the customer account will receive over-labels to place on affected inventory, with instructions for the customer staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. For distributors or those who have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0518-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology