FDA RECALL RECORD · Z-0519-2025

SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System Recall

SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System

FDA recall Class II

Ongoing

FDA lists this Class II recall for SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0519-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 28, 2024
FDA report date
Dec 4, 2024
Quantity in commerce
3 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System

Affected model & UDI checker

Search the exact identifiers published for Z-0519-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedSIGNA Architect AIR, Nuclear Magnetic Resonance Imaging SystemCheck source notice00195278283481
FDA code information

UDI/DI (01)00195278283481(11)231214(21)CNARC2300080TJ, System ID 82427110161; UDI/DI (01)00195278283481(11)230519(21)CNARC2300027TJ, System ID 82427090471; UDI/DI (01)00195278283481(11)231116(21)CNARC2300072TJ, System ID 82427120353; UDI/DI (01)00195278283481(11)230830(21)CNARC2300051TJ, System ID 82427140303; UDI/DI (01)00195278283481(11)230515(21)CNARC2300026TJ, System ID 82427200210; UDI/DI (01)00195278283481(11)230809(21)CNARC2300046TJ, System ID 82427090487; UDI/DI (01)00195278283481(11)230816(21)CNARC2300048TJ, System ID 82427010214; UDI/DI (01)00195278283481(11)230719(21)CNARC2300040TJ, System ID 82427070331; UDI/DI (01)00195278283481(11)230904(21)CNARC2300052TJ, System ID 82427090477; UDI/DI (01)00195278283481(11)231102(21)CNARC2300068TJ, System ID 82427160409; UDI/DI (01)00195278283481(11)230722(21)CNARC2300041TJ, System ID 82427040320; UDI/DI (01)00195278283481(11)231120(21)CNARC2300073TJ, System ID 82427090470; UDI/DI (01)00195278283481(11)231009(21)CNARC2300060TJ, System ID 82427030219; UDI/DI (01)00195278283481(11)240104(21)CNARC2400002TJ, System ID 82427090488; UDI/DI (01)00195278283481(11)231207(21)CNARC2300078TJ, System ID 82427090469; UDI/DI (01)00195278283481(11)230918(21)CNARC2300056TJ, System ID 82427200214; UDI/DI (01)00195278283481(11)240131(21)CNARC2400008TJ, System ID 82427030223; UDI/DI (01)00195278283481(11)240113(21)CNARC2400005TJ, System ID 82427060120; UDI/DI (01)00195278283481(11)231109(21)CNARC2300070TJ, System ID 82427130176; UDI/DI (01)00195278283481(11)230920(21)CNARC2300057TJ, System ID 82427100480; UDI/DI (01)00195278283481(11)231010(21)CNARC2300061TJ, System ID 82427100491; UDI/DI (01)00195278283481(11)230729(21)CNARC2300043TJ, System ID 82427250136; UDI/DI (01)00195278283481(11)240116(21)CNARC2400006TJ, System ID 82427250132; UDI/DI (01)00195278283481(11)240110(21)CNARC2400004TJ, System ID 82427110160; UDI/DI (01)00195278283481(11)231026(21)CNARC2300066TJ, System ID 82427100496; UDI/DI (01)00195278283481(11)231204(21)CNARC2300077TJ, System ID 82427100504; UDI/DI (01)00195278283481(11)230508(21)CNARC2300024TJ, System ID 82427160403; UDI/DI (01)00195278283481(11)231221(21)CNARC2300081TJ, System ID 82427030207; UDI/DI (01)00195278283481(11)240102(21)CNARC2400001TJ, System ID 82427120364; UDI/DI (01)00195278283481(11)231019(21)CNARC2300064TJ, System ID 82427120360; UDI/DI (01)00195278283481(11)240127(21)CNARC2400007TJ, System ID 82427200224; UDI/DI (01)00195278283481(11)231030(21)CNARC2300067TJ, System ID 82427200207; UDI/DI (01)00195278283481(11)231016(21)CNARC2300063TJ, System ID 82427140310; UDI/DI (01)00195278283481(11)230818(21)CNARC2300049TJ, System ID 82427140317; UDI/DI (01)00195278283481(11)230911(21)CNARC2300054TJ, System ID 82427140308; UDI/DI (01)00195278283481(11)230913(21)CNARC2300055TJ, System ID 82427190195

View the original FDA recall notice

Why was this product recalled?

Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.

What actions did GE Medical Systems, LLC take?

GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 10/28/20224 delivered through a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken for the continued safe usage of the affected devices: "Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." Please complete and return the attached acknowledgement form to [email protected]. For questions or concerns, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.