Brand Name: Z-800 Infusion System Product Name: Z-800WF Recall

Brand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision physician or other certified healthcare professional.

FDA recall Class III

Ongoing

FDA lists this Class III recall for Brand Name: Z-800 Infusion System Product Name: Z-800WF. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0523-2025
Recall classification
Class III
Enforcement status
Ongoing
Initiation date
Sep 30, 2024
FDA report date
Dec 4, 2024
Quantity in commerce
100 units
Distribution
See key details
FDA cause
Software design (manufacturing process)

What is being recalled?

Brand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision physician or other certified healthcare professional.

Affected model & UDI checker

Search the exact identifiers published for Z-0523-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Z-800WFBrand Name: Z-800 Infusion System Product Name: Z-800WFCheck source noticeNot specified
Not specifiedBrand Name: Z-800 Infusion System Product Name: Z-800WFCheck source notice00814371020006
FDA code information

Model/Catalog Number: Z-800WF; UDI-DI: 00814371020006; Serial Numbers: 500346, 500347, 500348, 500349, 500350, 500351, 500352, 500353, 500354, 500355, 500356, 500357, 500358, 500359, 500360, 500361, 500362, 500363, 500364, 500365, 500367, 500368, 500369, 500370, 500371, 500373, 500374, 500375, 500376, 500377, 500378, 500384, 500385, 500386, 500388, 500389, 500390, 500391, 500392, 500393, 500394, 502634, 503435, 503716, 503799, 503917, 503921, 504534, 504807, 504808, 504809, 504810, 504811, 504812, 504813, 504814, 504815, 504816, 504817, 504818, 504819, 504820, 504821, 504822, 504823, 504824, 504825, 504826, 504827, 504828, 504829, 504830, 504831, 504832, 504834, 504835, 504836, 504838, 505369, 505425, 505430, 505432, 505437, 505626, 505627, 505628, 505629, 505630, 505631, 505632, 505633, 505635, 505637, 505638, 505691, 505692, 505693, 506585, 506982, 506983.

View the original FDA recall notice

Why was this product recalled?

For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.

What actions did Zyno Medical LLC take?

On September 30, 2024 MEDICAL DEVICE NOTIFICATION letters were sent to customers. Mitigation: Zyno Medical is issuing this customer notification, advising customers that the expected User ID may be missing from completed therapies in the EMR Patient Record. Zyno Medical is planning to correct this issue through an update to the software. Recommended Actions: The user can locate the Patient Order in the Patient Record in the EMR and manually update the record to reflect the administrating clinician s User ID until the Z-800WF pump software version is updated. Follow the guidelines of your facility for performing manual updates in the EMR. Reporting Adverse Events: We strongly encourage healthcare providers and patients to promptly report any adverse events associated with the use of the Zyno Medical Z-800, Z-800F, Z-800W, and/or Z-800WF Infusion Pump. Please send your reports to Zyno Medical via email to [email protected]. Additionally, we urge you to report any negative side effects related to the Zyno Medical Z-800, Z-800F, Z-800W, and/or Z-800WF Infusion Pumps to the FDA. You can submit your report online at www.fda.gov/medwatch or by calling 1-800-FDA-1088. Your cooperation in reporting these events is vital for continuous safety monitoring of medical devices. For information about the Zyno Medical Z-800, Z-800F, Z-800W, and/or Z-800WF Infusion Pumps, including benefits and risks, refer to the Zyno Medical Z-800 Infusion Pump Instructions for Use (Z-800 800-IFU-6651, Z-800F 800-IFU-2602, Z-800W 800W-IFU-3614, and Z-800WF 800WF-IFU-4602) for complete description of the benefits and risks of using the Zyno Medical Z-800 family of pumps. Zyno Medical is committed to offering products which meet the highest quality standards and to continuous improvement in the interest of patient safety. We regret any inconvenience that this action may cause, but believe transparent communication will ensure you have timely, relevant information for managing your patients. If you

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology