Medline BUBBLE CPAP KIT, Medline Kit SKU HUD26007KIT Recall
Medline BUBBLE CPAP KIT, Medline Kit SKU HUD26007KIT
FDA recall Class II
Open , Classified
FDA lists this Class II recall for Medline BUBBLE CPAP KIT, Medline Kit SKU HUD26007KIT. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0523-2027
- Recall classification
- Class II
- FDA database status
- Open , Classified
- Initiation date
- Oct 7, 2026
- FDA website posting date
- Nov 20, 2026
- Quantity in commerce
- 24 units
- Distribution
- See key details
- FDA cause
- Not reported
What is being recalled?
Medline BUBBLE CPAP KIT, Medline Kit SKU HUD26007KIT
Affected model & UDI checker
Search the exact identifiers published for Z-0523-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Medline BUBBLE CPAP KIT, Medline Kit SKU HUD26007KIT | Check source notice | 40195327227198 |
FDA code information
UDI/DI 40195327227198 (each) 40195327227198 (case), Lot Number 25DBT748
Why was this product recalled?
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage.
What actions did Medline Industries, LP take?
Medline issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 10/07/2026 via USPS and email. The notice explained the issue, risk to health, and requested the affected product be destroyed. Credit will be issued once the response form is received. For questions, contact the Recall Department at 866-359-1704 or [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0523-2027
- FDA recall database source · Posted Nov 20, 2026 · Retrieved Nov 21, 2026
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology