smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP Recall
smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP
FDA recall Class II
Ongoing
FDA lists this Class II recall for smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0528-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 28, 2024
- FDA report date
- Dec 4, 2024
- Quantity in commerce
- 10,223 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP
Affected model & UDI checker
Search the exact identifiers published for Z-0528-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 14-504-68 | smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP | Check source notice | Not specified |
| 14-504-68JP | smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP | Check source notice | Not specified |
| Not specified | smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP | Check source notice | 35021239002586 |
FDA code information
REF 14-504-68, UDI/DI 35021239002586, Lot Numbers: 0003393, 0003394, 0003357, 0003358, 0003267, 0003368, 0003634, 0003254, 0003369, 0003165, 0003184, 0003196, 0003204, 0003206, 0003226, 0003256, 0003300, 0003307, 0003308, 0003314, 0003322, 0003359, 0003360, 0003361, 0003362, 0003363, 0003364, 0003377, 0003380, 0003635, 0003643; REF 14-504-68JP, UDI/DI 35021239002586, Lot Numbers: 0003369, 0003377, 0003393, 0003394, 0003634, 0003635, 0003643
Why was this product recalled?
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Smiths Medical ASD, Inc. take?
Smiths Medical issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 10/28/2024, mailed using a traceable means. The notice explained the issue, potential risk, and requested the consignee locate and destroy all affected products, and distribute the notice to all affected parties. Distributors were directed to notify their customers. For questions: About the communication: [email protected], 1-800-913-3897; To report adverse events: [email protected], 1-866-216-8806
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0528-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology