(1) Medline Pacer Implant Pack, REF DYNJ59194A, 4 kits/case, sterile; and (2) Medline Thoracotomy Morristown, REF DYNJ904220F, 1 kit/case, non-sterile. Recall
(1) Medline Pacer Implant Pack, REF DYNJ59194A, 4 kits/case, sterile; and (2) Medline Thoracotomy Morristown, REF DYNJ904220F, 1 kit/case, non-sterile.
FDA recall Class II
Ongoing
FDA lists this Class II recall for (1) Medline Pacer Implant Pack, REF DYNJ59194A, 4 kits/case, sterile; and (2) Medline Thoracotomy Morristown, REF DYNJ904220F, 1 kit/case, non-sterile.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0533-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 15, 2024
- FDA report date
- Dec 4, 2024
- Quantity in commerce
- 360 kits
- Distribution
- See key details
- FDA cause
- Mixed-up of materials/components
What is being recalled?
(1) Medline Pacer Implant Pack, REF DYNJ59194A, 4 kits/case, sterile; and (2) Medline Thoracotomy Morristown, REF DYNJ904220F, 1 kit/case, non-sterile.
Affected model & UDI checker
Search the exact identifiers published for Z-0533-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| DYNJ59194A | (1) Medline Pacer Implant Pack, REF DYNJ59194A, 4 kits/case, sterile; and (2) Medline Thoracotomy Morristown, REF DYNJ904220F, 1 kit/case, non-sterile. | Check source notice | Not specified |
| DYNJ904220F | (1) Medline Pacer Implant Pack, REF DYNJ59194A, 4 kits/case, sterile; and (2) Medline Thoracotomy Morristown, REF DYNJ904220F, 1 kit/case, non-sterile. | Check source notice | Not specified |
FDA code information
(1) REF DYNJ59194A - Lot numbers 24DLA916, exp. 12/31/2025; and 24BLA030, exp. 11/30/2025; UDI-DI each 10193489573480, UDI-DI case 40193489573481; and (2) REF DYNJ904220F - Lot number 24DLA070, exp. 11/30/2024, UDI-DI each 10195327362300, UDI-DI case 40195327362301.
Why was this product recalled?
Some of the suture packages within the convenience kits contain the incorrect needle type/size and suture diameter/length.
What actions did MEDLINE INDUSTRIES, LP - Northfield take?
The recalling firm issued letters on and dated 10/15/2024 via email and first-class mail explaining the issue, potential risk, and informing the consignee stickers are to be provided upon request to be placed onto the Medline pack instructing that the affected component is to be removed and replaced with an item in inventory. Required actions were included which were to (1) Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately; (2) Use the link and information in the letter to complete the response form. List the quantity of affected product in inventory on the form. Upon receipt of the submitted response form, the consignee will receive over-labels to place on affected inventory with instructions for the consignee staff to remove the affected component prior to using the kit. The labels will be provided via FedEx overnight; (4) If the consignee is a distributor or has resold or transferred this product to another company or individual, the consignee is to notify them of the recall communication.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0533-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology