FDA RECALL RECORD · Z-0534-2026

Brand Name: ReQuest Measles IgG Product Name: Rubeola IgG Enzyme Linked Immunoabsorbant Assay Recall

Brand Name: ReQuest Measles IgG Product Name: Rubeola IgG Enzyme Linked Immunoabsorbant Assay Model/Catalog Number: 01-190 Software Version: N/A Product Description: Measles IgG tests, performed using ELISA (Enzyme-Linked Immunosorbent Assay) method, used to detect the presence of IgG antibodies against measles virus in a patient's serum or plasma Component: not a component

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: ReQuest Measles IgG Product Name: Rubeola IgG Enzyme Linked Immunoabsorbant Assay. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0534-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 1, 2025
FDA report date
Nov 26, 2025
Quantity in commerce
Not reported
Distribution
See key details
FDA cause
No Marketing Application

What is being recalled?

Brand Name: ReQuest Measles IgG Product Name: Rubeola IgG Enzyme Linked Immunoabsorbant Assay Model/Catalog Number: 01-190 Software Version: N/A Product Description: Measles IgG tests, performed using ELISA (Enzyme-Linked Immunosorbent Assay) method, used to detect the presence of IgG antibodies against measles virus in a patient's serum or plasma Component: not a component

Affected model & UDI checker

Search the exact identifiers published for Z-0534-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
01-190Brand Name: ReQuest Measles IgG Product Name: Rubeola IgG Enzyme Linked Immunoabsorbant AssayCheck source noticeNot specified
FDA code information

Lot Code: DI# 00850487007357 Lots affected: L07024 and E03025

View the original FDA recall notice

Why was this product recalled?

Due to distributing the measles IgG IVD without a premarket approved/cleared.

What actions did Quest International, Inc. take?

On 10/01/2025, the firm email an "URGENT: Medical Device Recall Measles IgG IFU" letter to customers informing them that the Measles IgG Medical Device is subject to the limitations of exemption from premarket notification (510(k)) under 21 CFR866.9(c)(7) and is subject to the 510(k) premarket requirements prior to being marketed and distributed in the U.S. Customers are instructed to examine their inventory, discontinue use and quarantine the affected lot and review the updated Instruction for Use (IFU) and ensure all relevant personnel are informed of the changes. The previous version of the IFU should be replace with the updated version, Revision 9. The updated IFU (Revision 9) can also be requested by contacting Customer Support team at [email protected]. Awareness Technology, Inc. will be revising the intended use of the product to be only for qualitative detection of human IgG antibodies to measles virus in human serum by enzyme immunoassay. The company will also be removing language in the Instructions for Use (IFU) that could be interpreted as determining immune status, statements regarding paired sera for demonstrating seroconversion or significant rise in antibody levels, and any reference to semi-quantitative procedures. The updated IFU (Revision 9) will be included in all product shipments starting 10/01/2025. For questions or assistance, contact Customer Service at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology