Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm Recall
Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm Model/Catalog Number: 321-59-46 Software Version: N/A Product Description: rTSA Reamer, 46mm Component: Not a component
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0536-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 5, 2025
- FDA report date
- Nov 26, 2025
- Quantity in commerce
- 1
- Distribution
- See key details
- FDA cause
- Release of Material/Component prior to receiving test results
What is being recalled?
Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm Model/Catalog Number: 321-59-46 Software Version: N/A Product Description: rTSA Reamer, 46mm Component: Not a component
Affected model & UDI checker
Search the exact identifiers published for Z-0536-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 321-59-46 | Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm | Check source notice | Not specified |
| Not specified | Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm | Check source notice | 10885862605535 |
FDA code information
Lot Code: Part Number: 321-59-46, UDI: 10885862605535, Lot Number: 410398-014
Why was this product recalled?
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
What actions did Exactech, Inc. take?
On 09/05/2025, the firm emailed an "URGENT MEDICAL DEVICE RECALL NOTICE" to customers informing that, during routine inventory review, it was discovered that the a number of parts/products, which were pending design verification testing were incorrectly placed in active internal inventory rather than designated quarantine locations. Several part/products were distributed to customers. Customers are instructed to: "Carefully review this notification to ensure that they fully understand the issue and all actions required. "Immediately identify, quarantine, and return any of the subject devices in their inventory and/or customer s inventory. "Stop and cease use of any affected devices. "Notify all their accounts that may have received these products and extend this recall notice, including the Description of Issue and Potential Clinical Impact, to them. "Complete and Return the attached Recall Inventory Response Form to Exactech via email at [email protected]. For questions or assistance - contact Recall Support Team at [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0536-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology