OC-Auto FOBT-CHEK Negative Controls is the negative hemoglobin control materials designed to monitor FOBT (Fecal Occult Blood Test) Recall
OC-Auto FOBT-CHEK Negative Controls is the negative hemoglobin control materials designed to monitor FOBT (Fecal Occult Blood Test)
FDA recall Class III
Ongoing
FDA lists this Class III recall for OC-Auto FOBT-CHEK Negative Controls is the negative hemoglobin control materials designed to monitor FOBT (Fecal Occult Blood Test). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0538-2026
- Recall classification
- Class III
- Enforcement status
- Ongoing
- Initiation date
- Oct 6, 2025
- FDA report date
- Nov 26, 2025
- Quantity in commerce
- 403 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
OC-Auto FOBT-CHEK Negative Controls is the negative hemoglobin control materials designed to monitor FOBT (Fecal Occult Blood Test)
Affected model & UDI checker
Search the exact identifiers published for Z-0538-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | OC-Auto FOBT-CHEK Negative Controls is the negative hemoglobin control materials designed to monitor FOBT (Fecal Occult Blood Test) | Check source notice | 00855574005520 |
FDA code information
UDI: 00855574005520, Catalog Number(s): OCQN Lot Number(s): Lot 250805
Why was this product recalled?
A labeling error was identified on the affected product. The expiration date on the kit box label and on the vial labels incorrectly stated the expiration date as 2026-01-16. The correct expiration date should be 2027-01-17.
What actions did POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC take?
On October 6, 2025 URGENT PRODUCT RECALL: INCORRECT EXPIRATION DATE ON PRODUCT LABELING letters were sent to customers. ACTIONS REQUIRED BY YOU 1. Please trace the OCQN kits, lot 250805 in your inventory and at the customer sites listed in appendix one. If the kits have already been used no further action is required. There is no risk to patient care. 2. For customers with remaining stock there are two options:1) Manually correct the date on the packaging and vial labels to 2027-01-17 and add this notification to the kit box. OR 2) Request replacement stock with corrected labeling from Polymedco. ADDITIONAL INFORMATION Polymedco s inventory of Lot 250805 has been updated with new labels with the correct shelf life. All kits provided from September 25, 2025, will have the correct expiration date.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0538-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology