FDA RECALL RECORD · Z-0554-2025

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL Recall

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

FDA recall Class II

Ongoing

FDA lists this Class II recall for Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0554-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 9, 2024
FDA report date
Dec 4, 2024
Quantity in commerce
48
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Affected model & UDI checker

Search the exact identifiers published for Z-0554-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
SIL-40CAutosampler, Models: SIL-40C XR CL, SIL-40C X3 CLCheck source noticeNot specified
FDA code information

Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16.

View the original FDA recall notice

Why was this product recalled?

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Shimadzu Corp. Analytical & Measuring Instruments Div. take?

On 10/9/24, field safety notices were mailed to customers asking them to do the following: Check the Serial Numbers, and if the instruments are affected by this issue, contact your regional office. The corrective actions to be taken to affective products are (1) Add a cover part to the main board. (2) Update the SIL-40 firmware to Ver. 1.16 or above. For questions, please call 800-477-1227 between the hours of 8:30-5:00 EST or visit www.ssi.shimadzu.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology