Arial Server Software Versions 11.1.4 through 11.1.6, SKU 54305, in conjunction with the Arial Mobile Application or Arial Care Giver Application Recall
Arial Server Software Versions 11.1.4 through 11.1.6, SKU 54305, in conjunction with the Arial Mobile Application or Arial Care Giver Application
FDA recall Class II
Ongoing
FDA lists this Class II recall for Arial Server Software Versions 11.1.4 through 11.1.6, SKU 54305, in conjunction with the Arial Mobile Application or Arial Care Giver Application. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0570-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 1, 2024
- FDA report date
- Dec 4, 2024
- Quantity in commerce
- 111 systems
- Distribution
- See key details
- FDA cause
- Software Design Change
What is being recalled?
Arial Server Software Versions 11.1.4 through 11.1.6, SKU 54305, in conjunction with the Arial Mobile Application or Arial Care Giver Application
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
All Systems using Arial Server Software versions 11.1.4 through 11.1.6 in conjunction with the Arial Mobile Application or Arial Care Giver Application
Why was this product recalled?
Certain software versions will not notify end users of calls initiated by call points with the Arial Emergency and Nurse Call System.
What actions did Securitas Healthcare LLC take?
An URGENT: MEDICAL DEVICE CORRECTION letter dated 10/31/24 was sent to customers. Necessary Actions to be Taken by Clients Instructions for determining which versions of the Arial Server Software you are using are attached hereto as Exhibit C. Clients using Arial Server Software versions 11.1.4.9, 11.1.5.5, or 11.1.6.2 should: 1. Not decline previously cleared calls through the Arial App; 2. Restart the Arial Server Software if an end user has declined a previously completed call using the Arial App. After restarting the Arial Server Software, calls from the affected call point(s) will begin to be received by the Arial Server Software and the Arial Server Software will notify end users as intended. Instructions for restarting the Arial Server Software are included on Exhibit B. 3. Contact Securitas Healthcare Technical Support at 1 (800) 380-8883 to upgrade the Arial Server Software to version 11.1.7 or later. For general questions concerning this communication please contact Securitas Healthcare Technical Support at 1 (800) 380-8883. We ask that you please complete and return a copy of the acknowledgement form attached hereto as Exhibit A.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0570-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology