Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Recall
Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P Product Description: Greiner VACUETTE Blood Collection Tube with Clot Activator and Gel Separator and Standard Stopper (16 x 100 mm, 8ml)
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0571-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 19, 2025
- FDA report date
- Nov 26, 2025
- Quantity in commerce
- 1,980,000
- Distribution
- See key details
- FDA cause
- Labeling Change Control
What is being recalled?
Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P Product Description: Greiner VACUETTE Blood Collection Tube with Clot Activator and Gel Separator and Standard Stopper (16 x 100 mm, 8ml)
Affected model & UDI checker
Search the exact identifiers published for Z-0571-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 455071P | Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator | Check source notice | Not specified |
| Not specified | Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator | Check source notice | 29120017573061 |
FDA code information
Lot Code: Lot/Serial Numbers: Item# 455071P Lot# B250437N UDI-DI code: 29120017573061
Why was this product recalled?
Some of the Vacuette¿ CAT Serum Separator Clot Activator tubes were manufactured without gel separator, which could yield erroneous assay results and require the patient to return to the lab for recollection or repeat testing. This may result in delayed diagnosis, cause a delay in clinical decision-making or treatment and in some rare cases, yield incorrect results from less stable parameters.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Greiner Bio-One North America, Inc. take?
On 08/22/2025, the firm sent via email an "Urgent Medical Device Correction" letter to customer informing them that some Vacuette CAT Serum Separator Clot Activator tubes were manufactured without gel separator which may affect the stability of the sample. Customers are instructed to: " Stop use immediately: Discontinue use of the affected product (concerned Item/Lot) and isolate it within your facility. " Notify staff and customers: Inform all relevant employees and notify all customers to whom you distributed the affected products. " Assess risk: Evaluate potential risks or consequences in accordance with your internal procedures and take appropriate action. " Confirm product disposition: Complete the attached Product Disposition Site Confirmation Form and return it via fax to 877-217-7927 or email to [email protected]. Confirmation is required that all affected products have been discarded/destroyed. Replacement products will be provided upon receipt of your completed form. For questions, call Greiner Bio-One Technical Service at 800-515-8112.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0571-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology