Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0572-2025

TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509) Recall

TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509)

FDA recall Class II

Ongoing

FDA lists this Class II recall for TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0572-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 11, 2024
FDA report date
Dec 4, 2024
Quantity in commerce
160 units
Distribution
See key details
FDA cause
Process change control

What is being recalled?

TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509)

Affected model & UDI checker

Search the exact identifiers published for Z-0572-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5161-00-5048TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509)Check source noticeNot specified
Not specifiedTriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509)Check source notice00858325005688
FDA code information

Model No 5161-00-5048; UDI-DI 00858325005688; Lot Numbers: GB153869 and GB156110

View the original FDA recall notice

Why was this product recalled?

An out of specification component may result in the device exhibiting torque output higher than the design specification.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Gauthier Biomedical, Inc. take?

A Medical Device Recall (Removal) notification letter dated 11/11/24 was sent to the customer. Please Take the Following Steps: 1. Examine your inventory to determine if you have the affected lots and quarantine them immediately. DO NOT USE THE AFFECTED INSTRUMENTS. 2. Follow standard product return practices by contacting Heidi Wright at [email protected] to issue an RMA number and coordinate the return of the affected product(s) to the following address: ATTN: Heidi Wright Gauthier Biomedical, Inc. 2221 Washington Street Grafton, WI 53024 Note: If the Verification Form is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual on this page of the notification. 3. Review, complete, sign and return the attached Verification Form to Heidi Wright at [email protected] within five (5) business days of receipt of this notification. Please include in the email subject: CA 24-010. 4. Please complete the attached Verification Form even if you do not have the affected product lots on hand. 5. Forward this notice to anyone in your facility that needs to be informed (e.g., those who manage transport, store, stock, or use the affected products). 6. If any of the affected product lots have been forwarded to another facility, contact that facility and provide them with this notice. 7. Post a copy of this notice in a visible area for awareness and keep a copy for your records. 8. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA's MedWatch Adverse Event Reporting program online: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reportingprogram/reporting-serious-problems-fda. This medical device recall (removal) has been reported to the relevant health authorities. If you have any questions, please contact Heidi Wright, Quality Manager at Gauth

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology