Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0579-2025

Ziehm Vision RFD 3D. Interventinoal fluoroscopic x-ray system Recall

Ziehm Vision RFD 3D. Interventinoal fluoroscopic x-ray system

FDA recall Class II

Ongoing

FDA lists this Class II recall for Ziehm Vision RFD 3D. Interventinoal fluoroscopic x-ray system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0579-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 8, 2024
FDA report date
Dec 11, 2024
Quantity in commerce
50
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

Ziehm Vision RFD 3D. Interventinoal fluoroscopic x-ray system

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI: EZIEZIEHMVISIONRFD3D1. Serial Numbers: 21118, 21212, 22040, 22710, 22764, 23247, 23248, 23255, 23257, 23258, 23259, 23260, 23327, 23328, 23329, 23456, 23457, 23458, 23459, 23653, 23654, 23655, 23656, 23657, 23666, 23667, 23668, 23669, 23670, 23671, 23672, 23673, 23674, 23723, 23740, 23744, 23745, 23746, 23747, 23748, 23749, 23750, 23751, 23752, 23753, 23775, 23801, 23802, 23828, 23849.

View the original FDA recall notice

Why was this product recalled?

Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Orthoscan, Inc. take?

Ziehm Imaging, Inc. intends to notify the affected customers directly in the following manner: 1) Letter with return receipt requested, envelope, return, and reply to questionnaire, or E-mail contact. 2) Direct phone calls. Once contact is made and confirmed Ziehm Imaging Inc, shall provide replacement hand switch and instructions to the end user. Wired Hand switch 88036 Vl.02 will be requested to be returned by the customer/end user to Ziehm Imaging Inc. and will be subjected to be scrapped.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology