Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0581-2025

BIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor. Recall

BIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.

FDA recall Class II

Ongoing

FDA lists this Class II recall for BIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0581-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 29, 2024
FDA report date
Dec 11, 2024
Quantity in commerce
2,156 units
Distribution
See key details
FDA cause
Package design/selection

What is being recalled?

BIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.

Affected model & UDI checker

Search the exact identifiers published for Z-0581-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedBIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.Check source notice03596010595072
Not specifiedBIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.Check source notice03596010595089
Not specifiedBIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.Check source notice03596010656452
Not specifiedBIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.Check source notice03596010656469
FDA code information

Part No. 72201541, 72201542, 72203280, 72203281; UDI-DI: 03596010595072, 03596010595089, 03596010656452, 03596010656469; Batch No. 2146369, 2151242, 2151690, 2154691, 2153489, 2151692.

View the original FDA recall notice

Why was this product recalled?

Sterile barrier breach due to inadequate packaging design that could not hold the geometry of the device.

What actions did Smith & Nephew, Inc. take?

An URGENT MEDICAL DEVICE RECALL NOTICE dated 10/29/24 was mailed to consignees. This notification instructed consignees to inspect their inventory for affected devices and quarantine them immediately. If product was further distributed consignees are to notify their customers of this recall and ensure recall actions are completed. Consignees are to complete the provided response form to facilitate the return of devices and email the completed form to [email protected]. Consignees who do not have product on hand are to complete and return the recall response form as well. Any questions about this recall are to be emailed to [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology