Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0582-2025

Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile. Recall

Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0582-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 31, 2024
FDA report date
Dec 11, 2024
Quantity in commerce
46 devices
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.

Affected model & UDI checker

Search the exact identifiers published for Z-0582-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
G24663Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.Check source noticeNot specified
Not specifiedCook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.Check source notice00827002246638
FDA code information

UDI-DI 00827002246638, Lot numbers W4844245, exp. 6/10/2027; W4847501, exp. 6/20/2027; and W4851589, exp. 7/8/2027.

View the original FDA recall notice

Why was this product recalled?

Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.

What actions did Wilson-Cook Medical Inc. take?

The recalling firm issued letters on 10/31/2024 dated 11/1/2024 via email to their consignees extending the voluntary recall from 29 August 2024 to include additional lot numbers of Hemospray Endoscopic Hemostat devices due to the receipt of additional complaints and adverse events since the recall was originally initiated. The consignee was instructed to review the attached list to ensure that affected devices are quarantined and returned to the recalling firm. The letter goes onto discuss the indications for use for the 2 previously recalled devices (HEMO-7 and HEMO-10) and provides the indications for use for the HEMO-7-EU device, which were different from the HEMO-7 and HEMO-10 devices. The reason for recall was explained, the risk to health, and specific product identification and device information (device brand name, reference part number, model number, and UDI-DI number). The actions to be taken by the customer were provided, which included immediately examining inventory and quarantining affected devices, return of the devices, completing the Acknowledgement and Receipt Form within 5 business days of receiving the letter, sharing the notice with appropriate personnel, including down to the user level with or outside their organization to anywhere the devices were transferred, and immediately reporting adverse events to Cook Medical Customer Relations. Actions to be taken by the distributor included the same details as for the customer except that the distributor was to notify any organizations that have been distributed the devices. On 11/20/2024, the firm provided a copy of the recall letter for the HEMO-7-EU dated 11/20/2024 they issued via email to their consignees. The letter explained the HEMO-7-EU was identified as also needing to be recalled when the firm expanded their recall of HEMO-7 and HEMO-10. The letter provided the reason for recall, risk to health, product identification and device information, and actions to be taken by the distri

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology