Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0583-2025

Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344. Recall

Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0583-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 16, 2024
FDA report date
Dec 11, 2024
Quantity in commerce
301 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.

Affected model & UDI checker

Search the exact identifiers published for Z-0583-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
728333Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.Check source noticeNot specified
728340Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.Check source noticeNot specified
728344Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.Check source noticeNot specified
Not specifiedSpectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.Check source notice00884838101111
FDA code information

UDI-DI: (01)00884838101111. Software Version: 5.0.X. Model Number (Serial Numbers): 728333 (10266, 10261, 10264, 10219, 10265, 10197, 10182, 10175, 10192, 10179, 10196, 10180, 10188, 10190, 10181, 10183, 10178, 10186, 10185, 10199, 10184, 10198, 10193, 10187, 10189, 10195, 10191, 10194, 10018, 10023, 10027, 10062, 10072, 10078, 10079, 10080, 10084, 10088, 10101, 10104, 10112, 10117, 10113, 10116, 10032, 10049, 10133, 10140, 10152, 10153, 10166, 10170, 10171, 10173, 10172, 10174, 10016, 10015, 10017, 10019, 10022, 10021, 10020, 10025, 10024, 10033, 10031, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10043, 10042, 10014, 10044, 10046, 10048, 10045, 10047, 10051, 10050, 10053, 10055, 10052, 10056, 10058, 10057, 10059, 10060, 10061, 10065, 10066, 10068, 10067, 10070, 10071, 10063, 10069, 10074, 10075, 10077, 10076, 10081, 10082, 10087, 10089, 10085, 10086, 10091, 10094, 10093, 10095, 10096, 10090, 10099, 10098, 10102, 10100, 10103, 10097, 10105, 10106, 10107, 10108, 10109, 396001, 396003, 10110, 10111, 10114, 10115, 10122, 10118, 10120, 10121, 10123, 10119, 10124, 10125, 10126, 10127, 10129, 10128, 10131, 10130, 10132, 10135, 10136, 10134, 10137, 10139, 10138, 10142, 10141, 10143, 10144, 10145, 10146, 10147, 10149, 10150, 10151, 10155, 10156, 10159, 10148, 10160, 10161, 10162, 10158, 10164, 10163, 10165, 10167, 10168, 10169, 10013, 10177, 10176, 10210, 10218, 10217, 10206, 10202, 10200, 10211, 10207, 10204, 396004, 10208, 10201, 10205, 10203, 10212, 10213, 10225, 10214, 10220, 10215, 10222, 10227, 10247, 10209, 10226, 10230, 10228, 10232, 10231, 10229, 10221, 10233, 10234, 10235, 10237, 10238, 10223, 10239, 10241, 10242, 10236, 10245, 396005, 10248, 10249, 10243, 10244, 10250, 10259, 10258, 10157, 10262, 10246, 10255, 10254, 10240, 10251, 10256, 10224, 10253, 10257, 10260, 10263), 728340 (397060, 397061, 397022, 397023, 397020, 397024, 397021, 397025, 397027, 397031, 397026, 397033, 397032, 397030, 397028, 397029, 397015, 397016, 397004, 397007, 397005, 397002, 397006, 397008, 397010, 397003, 397011, 397013, 397009, 397012, 397014, 397017, 397019, 397018, 397036, 397037, 397038, 397034, 397041, 397040, 397039, 397044, 397046, 397043, 397045, 397047, 397048, 397049, 397053, 397050, 397054, 397055, 397056, 397058, 397059), 728344 (397057).

View the original FDA recall notice

Why was this product recalled?

Potential for a plan box not updated issue during Interventional procedure with Spectral CT that may lead to a collision with the operator or with the needle placed inside the patient, and an incorrect patient ID software issue that may lead to misdiagnosis.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips North America Llc take?

Philips notified consignees on about 11/16/2024, via certified mail. Consignees were instructed to identify any affected units on hand, identify if those units contain the affected software version, complete and return the response form, and circulate the Urgent Field Safety Notice Letter to all users of the device. To protect the user and patient from the issue of Plan Box not Updated, remain vigilant and confirm table position during procedures. Press any of the Emergency STOP buttons to stop unintended patient table motion and follow safety instructions in Section 6 of the IFU: Working with Specialized Exams. To avoid the issue of Wrong Patient ID with Anonymous patient, the operator should verify the patient information loaded into the patient data field is correct before scanning and rename the Anonymous record with the patient s name at the time of the scan.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.