Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0587-2026

Quantum Intersegmental Table Recall

Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table

FDA recall Class II

Ongoing

FDA lists this Class II recall for Quantum Intersegmental Table. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0587-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 9, 2025
FDA report date
Dec 3, 2025
Quantity in commerce
34 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table

Affected model & UDI checker

Search the exact identifiers published for Z-0587-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Q400Quantum Intersegmental TableCheck source noticeNot specified
FDA code information

All serial numbers.

View the original FDA recall notice

Why was this product recalled?

The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.

What actions did Pivotal Health Solutions, Inc. take?

Pivotal Health Solutions issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 10/9/2025 via certified mail and Fed Ex. The notice explained the issue with the device, risk to health and requested the following: Immediately stop using table(s) with heat. The heat function must be disabled. You have been identified as having been sold a Quantum Table with heat. The serial numbers of the tables impacted are below. If you are unsure if your table has a heat function, please identify the serial number on the Quantum table(s), located on the lower section of table by the power cable inlet, and verify the serial number is on the list. The notice also supplied instructions for disconnecting and disabling the heat function o the Q400 tables. Contact information for questions: o Matt White ([email protected]) or Julee Driver ([email protected]), Quality Dept, at 605-753-0110 o Brenda Anderson ([email protected]), Service Dept at 605-753-0110

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology