Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated Recall
Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated
FDA recall Class II
Ongoing
FDA lists this Class II recall for Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0589-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 12, 2025
- FDA report date
- Dec 3, 2025
- Quantity in commerce
- Total devices 1821 (US=1794 and O.U.S.=27) units
- Distribution
- See key details
- FDA cause
- Equipment maintenance
What is being recalled?
Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated
Affected model & UDI checker
Search the exact identifiers published for Z-0589-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 352506070E | Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated | Check source notice | Not specified |
| Not specified | Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated | Check source notice | 00886333217151 |
FDA code information
Model Number: 352506070E UDI-DI code: 00886333217151 Lot Numbers: 11633166 11633125 11629543 11629273 11628836
Why was this product recalled?
Due to complaints of increased resistance when advancing the dilator within the introducer sheath. This issue could potential result in procedural delays and/or risk of venous intimal injuries if excessive force is applied
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Argon Medical Devices, Inc. take?
On 09/12/2025, the firm sent via certified mail an "Urgent - Product Recall Notice" to customers. On 9/2025 the firm sent a follow-up "URGENT - MEDICAL DEVICE RECALL NOTICE" dated 9/29/2025 informing them that Argon Medical Devices is voluntarily recalling five affected lots of the Option ELITE Vena Cava Filter System products in response to reports of increased resistance when advancing the dilator within the introducer sheath. Customers are instructed to: 1. Immediately discontinue use of the affected devices listed in the Recall Letter. 2. Segregate and quarantine the affected devices. 3. Complete and return the response form included at the end of this notification via regular mail or email by 9/26/2025. 4. Return all affected devices to Argon Medical Devices using the instructions provided below. 5. Share this recall notice with the following parties to ensure consignees receive a copy and respond accordingly: a. Anyone within their organization who may use this device. b. Any organization where the potentially affected products have been transferred or further distributed. c. Anyone who needs to be aware within their organization. The product may be returned at Argon s expense to the address below with the returned goods authorization (RGA# 29074) clearly marked. RGA# 29074 Argon Medical Devices, Inc. 1445 Flat Creek Road Athens, TX 75751 USA For any questions contact/email [email protected], or [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0589-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026