Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0590-2026

ErgoStar CM 40 Recall

ErgoStar CM 40 Model/Catalog Number: MP01840 Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal tube mask) of a mechanically ventilated patient.

FDA recall Class I

Ongoing

FDA lists this Class I recall for ErgoStar CM 40. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0590-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Nov 10, 2025
FDA report date
Dec 17, 2025
Quantity in commerce
14420 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

ErgoStar CM 40 Model/Catalog Number: MP01840 Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal tube mask) of a mechanically ventilated patient.

Affected model & UDI checker

Search the exact identifiers published for Z-0590-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MP01840ErgoStar CM 40Check source noticeNot specified
Not specifiedErgoStar CM 40Check source notice04048675420736
FDA code information

Model/Catalog Number: MP01840; UDI Number:04048675420736; Lot numbers: All lot numbers

View the original FDA recall notice

Why was this product recalled?

Multiple complaints were reported in which cracks formed in the hose of specific catheter mounts.

What actions did Draeger, Inc. take?

On November 10, 2025, Urgent Medical Device Recall letters were sent to customers. Necessary Action: Identify if you have any affected product. Do not use any unused ErgoStar CM 40/CM 45/CM 55/CM 60 in your stock. Unused stock should be removed and returned to Draeger. Draeger recommends the alternative product MP01850 ErgoStar CM 50 as a suitable replacement. Please ensure that all potential users in your facility are made aware of this Urgent Medical Device Recall Notice and the information contained herein. If you have forwarded affected products to third parties, please also forward this Urgent Medical Device Recall notification.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology