Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0594-2026

CareLink Clinic, REF: MMT-7350 Recall

CareLink Clinic, REF: MMT-7350

FDA recall Class II

Ongoing

FDA lists this Class II recall for CareLink Clinic, REF: MMT-7350. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0594-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 21, 2025
FDA report date
Dec 3, 2025
Quantity in commerce
12126
Distribution
See key details
FDA cause
Software change control

What is being recalled?

CareLink Clinic, REF: MMT-7350

Affected model & UDI checker

Search the exact identifiers published for Z-0594-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MMT-7350CareLink Clinic, REF: MMT-7350Check source noticeNot specified
FDA code information

UDI-DI: 0763000B00008748K. Software Versions: 4.2B Software Available: US: Between September 4th, 2025, and October 3rd, 2025. OUS: Between September 11th, 2025, and October 4th, 2025. US (Clinical): Between August 21st, 2025, and October 3rd, 2025. OUS (Clinical): Between August 21st, 2025, and October 4th, 2025.

View the original FDA recall notice

Why was this product recalled?

Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview graph may lead to hypoglycemia or hyperglycemia.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medtronic MiniMed, Inc. take?

On 10/21/2025, "URGENT MEDICAL DEVICE CORRECTION" notices were emailed to customers informing them of the following: software version 4.2C was released to correct the issues with the data displayed on 24-hour Sensor Glucose Overview graph. If you made individual diabetes treatment recommendations to patients based on the hypoglycemic patterns or on the 30-day sensor glucose data displayed on the 24-hour Sensor Glucose Overview graph with affected software, revisit those recommendations utilizing patients individual CareLink reports, as they were not impacted by this issue. Complete and return the confirmation form: https://info.medtronicdiabetes.com/carelink-clinic-dashboard-hcp or email form to [email protected] For questions, contact the firm's 24-Hour Technical Support line at 1-800-646-4633, option 9.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology