GE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as: Recall
GE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as:
FDA recall Class II
Ongoing
FDA lists this Class II recall for GE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as:. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0600-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 5, 2024
- FDA report date
- Dec 11, 2024
- Quantity in commerce
- 11 units
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
GE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as:
Affected model & UDI checker
Search the exact identifiers published for Z-0600-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 2374681-17 | GE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as: | Check source notice | Not specified |
| Not specified | GE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as: | Check source notice | 00840682118699 |
FDA code information
UDI/DI 00840682118699, Serial/Sales Order Numbers: CBDYG2400058HM, CBDYG2400057HM, CBDYG2400056HM, CBDYG2400051HM, CBDYG2400052HM, CBDYG2400040HM, CBDYG2400053HM, CBDYG2400060HM, CBDYG2400054HM, CBDYG2400055HM, CBDYG2400059HM.
Why was this product recalled?
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did GE Healthcare (China) Co., Ltd. take?
GE HealthCare issued an URGENT MEDICAL DEVICE notice to its consignees on 11/05/2024 via mailing using a traceable means. The notice explained the issue, potential safety issue, and requested the following actions be taken: " You can continue to use your system. When utilizing helical, cine, and cardiac scan modes, review the image data set for any potential rotation. When using these modes during interventional procedures or radiation therapy planning, adhere to standard clinical quality checks and workflows. If image rotation is detected, consult with the physician to determine whether a rescan is necessary or if the images need to be corrected during post-processing to ensure accurate interpretation and planning. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions." A GE HealthCare representative will contact the consignee to arrange for the correction. For questions or concerns: please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0600-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology