EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Recall
EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
FDA recall Class II
Ongoing
FDA lists this Class II recall for EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0620-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 31, 2024
- FDA report date
- Dec 11, 2024
- Quantity in commerce
- 1,770 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
Affected model & UDI checker
Search the exact identifiers published for Z-0620-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| GIF-1TH190 | EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. | Check source notice | 04953170343360 |
| Not specified | EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. | Check source notice | 04953170437014 |
FDA code information
Model Number: GIF-1TH190; UDI: 04953170343360, 04953170437014; Serial Number Range: Before 2446103 or After 2500453
Why was this product recalled?
High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).
What actions did Olympus Corporation of the Americas take?
Olympus notified consignees via letter on about 10/31/2024. Consignees were informed of the issue and of the upcoming revision of the labeling such that the GIF-1TH190 is no longer a compatible endoscope for reprocessing with Acecide-C in the OER-Pro. They were instructed that they would be contacted by their Olympus sales representative within the month to discuss potential OER-Elite and GIF-1TH190 transition programs. Additionally, they were instructed to carefully read the notification, ensure reprocessing staff and other users of the OER-Pro AER are aware of the corrective action, acknowledge receipt of the notification through the Olympus web portal, and to identify and notify recipients of the affected devices if they were further distributed.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0620-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology