Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0620-2025

EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Recall

EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190

FDA recall Class II

Ongoing

FDA lists this Class II recall for EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0620-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 31, 2024
FDA report date
Dec 11, 2024
Quantity in commerce
1,770 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190

Affected model & UDI checker

Search the exact identifiers published for Z-0620-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
GIF-1TH190EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190.Check source notice04953170343360
Not specifiedEVIS EXERA III Gastrointestinal Videoscope GIF-1TH190.Check source notice04953170437014
FDA code information

Model Number: GIF-1TH190; UDI: 04953170343360, 04953170437014; Serial Number Range: Before 2446103 or After 2500453

View the original FDA recall notice

Why was this product recalled?

High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).

What actions did Olympus Corporation of the Americas take?

Olympus notified consignees via letter on about 10/31/2024. Consignees were informed of the issue and of the upcoming revision of the labeling such that the GIF-1TH190 is no longer a compatible endoscope for reprocessing with Acecide-C in the OER-Pro. They were instructed that they would be contacted by their Olympus sales representative within the month to discuss potential OER-Elite and GIF-1TH190 transition programs. Additionally, they were instructed to carefully read the notification, ensure reprocessing staff and other users of the OER-Pro AER are aware of the corrective action, acknowledge receipt of the notification through the Olympus web portal, and to identify and notify recipients of the affected devices if they were further distributed.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology