ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The Chemolock is needle free and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel. Recall
ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The Chemolock is needle free and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.
FDA recall Class II
Ongoing
FDA lists this Class II recall for ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The Chemolock is needle free and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0652-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 6, 2024
- FDA report date
- Dec 18, 2024
- Quantity in commerce
- 6000 units
- Distribution
- See key details
- FDA cause
- Packaging process control
What is being recalled?
ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The Chemolock is needle free and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.
Affected model & UDI checker
Search the exact identifiers published for Z-0652-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| CL-80 | ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The Chemolock is needle free and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel. | Check source notice | Not specified |
| Not specified | ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The Chemolock is needle free and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel. | Check source notice | 10887709057125 |
FDA code information
Model/Catalog Number: CL-80 UDI code: (01)10887709057125(17)290401(30)50(10)13959938 Lot Number: 13959938
Why was this product recalled?
Due to finished goods incorrectly packaged and labeled.
What actions did Icu Medical, Inc. take?
On 11/06/2024, the firm send an "URGENT Medical Device Correction" Letter via third party (Sedgwick, Inc.) to customers informing them that product was incorrectly packaged and labeled. Product CL-80 package incorrectly contained product code CL-80S. The CL-80 is a used with a 20 mm or a 28 mm vial where as CL-80S is only use for a 20 mm vial. The issue is using the incorrect adapter may result in leakages. Actions to be taken by the Customer: Discontinue use mislabeled lot 13953938 of CL-80 and discard remaining product from this specific lot. 1. Check all inventory locations within your institution for the affected product listed in the notification and discontinue use. Quarantine and destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2. Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there./// 3. Complete and return the attached Customer Response Form to [email protected] within 10 days of receipt to acknowledge your understanding of this notification./ / 4. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]. For questions/assistance, contact: Global Complaint Management 1-844-654-7780 (M-F, 8:00am 5:00pm CT) or [email protected] To report adverse events or product complaints Customer Service 1-866-829-9025, option 8 or [email protected] (M-F, 8:00 am 6:00 pm CT) Additional information or to request a credit. Field Action Processing 1-888-843-0245 [email protected] Questions about this communication
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0652-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.