MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71 Recall
MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
FDA recall Class II
Ongoing
FDA lists this Class II recall for MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0665-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 28, 2023
- FDA report date
- Dec 18, 2024
- Quantity in commerce
- 18
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
Affected model & UDI checker
Search the exact identifiers published for Z-0665-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71 | Check source notice | 9357123000037 |
| Not specified | MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71 | Check source notice | 935712300005 |
FDA code information
MXU-RV35 UDI-DI: 9357123000037 and MXU-RV71 UDI-DI: 935712300005 using Control Board PCBA, 12425-03 (or earlier). Serial Numbers: MSN00334, MSN00348, MSN00349, MSN00351, MSN00352, MSN00361, MSN00364, MSN00365,MSN00382, MSN00383, MSN00384, MSN00385, MSN00386, MSN00387, MSN00388, MSN00390, MSN00391, MSN00395
Why was this product recalled?
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
What actions did Micro-X Ltd. take?
On 11/28/23 Modification 17 was sent to customers telling them that Control Boards need to be replaced with Control Board PCBA, 12425-04 that will eliminate the issue. For technical assistance contact: [email protected], 1-206-249-8764 Firm asked distributors to do the following: On next visit please confirm that the bulletins were completed and return the documentation.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0665-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.