Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0666-2025

MICRO-X Rover Mobile X-ray System, # MXU-RV19 Recall

MICRO-X Rover Mobile X-ray System, # MXU-RV19

FDA recall Class II

Ongoing

FDA lists this Class II recall for MICRO-X Rover Mobile X-ray System, # MXU-RV19. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0666-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 23, 2024
FDA report date
Dec 18, 2024
Quantity in commerce
23
Distribution
See key details
FDA cause
Device Design

What is being recalled?

MICRO-X Rover Mobile X-ray System, # MXU-RV19

Affected model & UDI checker

Search the exact identifiers published for Z-0666-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMICRO-X Rover Mobile X-ray System, # MXU-RV19Check source notice09357123000013
FDA code information

UDI-DI: 09357123000013. Serial Numbers: 262, 274, 320, 329, 237, 269, 271, 290, 292, 307, 310, 311, 312, 321, 322, 323, 327, 330, 331, 332, 333, 234, 266.

View the original FDA recall notice

Why was this product recalled?

During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.

What actions did Micro-X Ltd. take?

Starting on 10/23/24 Service Bulletin SB002-22 was sent to customers informing them that a field upgrade was released for the installation of an exposure switch holder. Customers were asked to complete and return an acknowledgement response form to [email protected] For further information, contact your local service representative at [email protected], 1-206-249-8764.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology