Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0668-2025

Brand Name: Atlan Product Name: Atlan A300XL Recall

Brand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Brand Name: Atlan Product Name: Atlan A300XL. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0668-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Oct 15, 2024
FDA report date
Dec 25, 2024
Quantity in commerce
25 units (OUS Only)
Distribution
See key details
FDA cause
Process control

What is being recalled?

Brand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.

Affected model & UDI checker

Search the exact identifiers published for Z-0668-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
8621400Brand Name: Atlan Product Name: Atlan A300XLCheck source noticeNot specified
Not specifiedBrand Name: Atlan Product Name: Atlan A300XLCheck source notice04048675556343
FDA code information

Model/Catalog Number: 8621400; UDI-DI Number: 04048675556343; Serial Numbers shipped directly from Draegerwerk AG & Co., KGaA to OUS: ASTB-0051; 0053; 0054; 0055; 0063; 0065; 0093; 0162 ASTE-0219; 0220; 0221; 0222; 0377 ASTF-0472; 0473 ASTH-0005; 0006; 0026; 0027; 0047; 0186; 0223; 0224; 0401 ASTJ-0170

View the original FDA recall notice

Why was this product recalled?

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

What actions did Draeger Medical, Inc. take?

On October 15, 2024 Urgent Medical Device Correction letters were sent to customers. Necessary Action: The corrective action is to replace the ventilator motor assembly. You will be contacted by your local Draeger Service Representative to arrange a date for the ventilator motor assembly replacement. Until the corrective action is implemented, you may continue operating your devices with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Monitor the patient s condition specifically their oxygenation status as a brief cessation of ventilation can potentially lead to e.g.; hypoxia, loss of lung recruitment, bradycardia and cardiac arrest. If preferred, the alarm priority of the "Ventilator error!!!" can be downgraded with "ALARM RESET" after switching to Man/Spont ventilation mode. Please ensure that all users of the Draeger Atlan as well as other persons concerned within your organization are made aware of this Urgent Medical Device Correction notice. Complete and return the Medical Device Correction Return Response Acknowledgment and Receipt Form to confirm that you have received this information. If you have made the products available to third parties, please forward this Urgent Medical Device Correction notice to them. If you have any questions regarding the operation of your Atlan A350/A350XL anesthesia workstation, please contact Draeger Service Technical Support 8AM-8PM EST at 1-800-437-2437 (press 2, then 2, then 2 again). Questions regarding this Urgent Medical Device Correction notice, contact Michael Kelhart 8AM-4:30PM EST at 267-664-1131 or via email at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology