FDA RECALL RECORD · Z-0669-2025

Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Recall

Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Number: F0101

FDA recall Class II

Completed

FDA lists this Class II recall for Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0669-2025
Recall classification
Class II
Enforcement status
Completed
Initiation date
Oct 10, 2022
FDA report date
Dec 18, 2024
Quantity in commerce
178 units
Distribution
See key details
FDA cause
Labeling Change Control

What is being recalled?

Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Number: F0101

Affected model & UDI checker

Search the exact identifiers published for Z-0669-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
F0101Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guideCheck source noticeNot specified
Not specifiedHologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guideCheck source notice15420045515390
FDA code information

Model No F0101. UDI 15420045515390. Lot Numbers: 186728, SS1-200727, SS1-181121, SS1-190301, SS1-190501, 175517, 175516, 162427, 140718-01, 171056, 175515.

View the original FDA recall notice

Why was this product recalled?

IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set

What actions did Hologic, Inc. take?

Hologic issued Letter-Revised Sterilization Instructions for Sizer Set on 10/10/22 via Fed'X. The letter states reason for recall, health risk and action to take: Effective immediately, please steam autoclave F0101 in accordance with such IFU, subject to the following parameters. Any additional parameters contained in Hologic s IFU for F0101 should no longer be utilized. link to their Instructions for Use (IFU) below: https://summitmedicalusa.com/wordpress/wp-content/uploads/2021/05/40192-01_RevZ_IS-IFU_Multilingual.pdf Please also note that when sterilizing F0101, one (1) layer of wrapping material (or one layer of 1 ply wrap) is necessary to provide a sterile barrier for each sterilization tray. Please do not utilize 2 layers or one layer of 2-ply wrap. In the future, Hologic will be updating the IFU supplied with the sizers within F0101, to reflect this change. mailto:[email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology