Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Recall
Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Number: F0101
FDA recall Class II
Completed
FDA lists this Class II recall for Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0669-2025
- Recall classification
- Class II
- Enforcement status
- Completed
- Initiation date
- Oct 10, 2022
- FDA report date
- Dec 18, 2024
- Quantity in commerce
- 178 units
- Distribution
- See key details
- FDA cause
- Labeling Change Control
What is being recalled?
Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Number: F0101
Affected model & UDI checker
Search the exact identifiers published for Z-0669-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| F0101 | Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide | Check source notice | Not specified |
| Not specified | Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide | Check source notice | 15420045515390 |
FDA code information
Model No F0101. UDI 15420045515390. Lot Numbers: 186728, SS1-200727, SS1-181121, SS1-190301, SS1-190501, 175517, 175516, 162427, 140718-01, 171056, 175515.
Why was this product recalled?
IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set
What actions did Hologic, Inc. take?
Hologic issued Letter-Revised Sterilization Instructions for Sizer Set on 10/10/22 via Fed'X. The letter states reason for recall, health risk and action to take: Effective immediately, please steam autoclave F0101 in accordance with such IFU, subject to the following parameters. Any additional parameters contained in Hologic s IFU for F0101 should no longer be utilized. link to their Instructions for Use (IFU) below: https://summitmedicalusa.com/wordpress/wp-content/uploads/2021/05/40192-01_RevZ_IS-IFU_Multilingual.pdf Please also note that when sterilizing F0101, one (1) layer of wrapping material (or one layer of 1 ply wrap) is necessary to provide a sterile barrier for each sterilization tray. Please do not utilize 2 layers or one layer of 2-ply wrap. In the future, Hologic will be updating the IFU supplied with the sizers within F0101, to reflect this change. mailto:[email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0669-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology