Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma. Recall
Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0673-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 8, 2024
- FDA report date
- Dec 18, 2024
- Quantity in commerce
- 31 US; 167 OUS
- Distribution
- See key details
- FDA cause
- Process design
What is being recalled?
Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.
Affected model & UDI checker
Search the exact identifiers published for Z-0673-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| A16972 | Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma. | Check source notice | Not specified |
| Not specified | Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma. | Check source notice | 15099590201937 |
FDA code information
UDI-DI 15099590201937 Lots 338921 339071 372223 439138 439205 439206 439207 439650 439801 439889 440026 440090 472012 472092 472136 472154
Why was this product recalled?
The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.
What actions did BECKMAN COULTER, INC. take?
On November 8, 2024, the firm notified affected customers via URGENT MEDICAL DEVICE RECALL letters that the affected assay may produce falsely decreased results if EDTA and lithium heparin plasma samples are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the test. Access Intact PTH assay Routine Mode and Intraoperative Mode users can select any of the following options: . Use serum as the sample type. . Perform testing on an alternate methodology, such as the Access 2 or UniCel DxI 600 or 800 Immunoassay Analyzers. . Continue to use EDTA and/or lithium heparin plasma sample types if the tests are completed within the timeframes listed below upon sample presentation to the DxI 9000 Access Immunoassay Analyzer (see page 3 for instructions): o Access Intact PTH assay Routine Mode: 25 minutes. o Access Intact PTH assay Intraoperative Mode: 20 minutes. . Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to perform a retrospective review of the patient results generated from Access Intact PTH Routine Mode plasma samples on the DxI 9000 to determine if retesting is necessary.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0673-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology