Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10 Recall
Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper FD10 OR Table, Model Number: 722022. 6. Allura Xper FD20 OR Table, Model Number: 722023. 7. Allura Xper FD20 Biplane OR Table, Model Number: 722025.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0676-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 11, 2024
- FDA report date
- Dec 18, 2024
- Quantity in commerce
- 2,782 units
- Distribution
- See key details
- FDA cause
- Equipment maintenance
What is being recalled?
Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper FD10 OR Table, Model Number: 722022. 6. Allura Xper FD20 OR Table, Model Number: 722023. 7. Allura Xper FD20 Biplane OR Table, Model Number: 722025.
Affected model & UDI checker
Search the exact identifiers published for Z-0676-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 722010 | Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10 | Check source notice | Not specified |
| 722012 | Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10 | Check source notice | Not specified |
| 722011 | Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10 | Check source notice | Not specified |
| 722013 | Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10 | Check source notice | Not specified |
| 722022 | Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10 | Check source notice | Not specified |
| 722023 | Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10 | Check source notice | Not specified |
| 722025 | Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10 | Check source notice | Not specified |
FDA code information
1. Model Number: 722010. UDI (Serial Numbers): (01)00884838054189(21)164 (164); (01)00884838059030(21)272 (272); (01)00884838059030(21)819 (819); (01)00884838059030(21)828 (828); No UDI (33, 39, 42, 43, 65, 67, 69, 79, 107, 119, 127, 158, 160, 169, 177, 190, 191, 209, 213, 214, 215, 218, 227, 229, 239, 240, 241, 252, 274, 280, 285, 286, 290, 296, 298, 306, 308, 316, 342, 343, 344, 360, 364, 365, 395, 411, 414, 422, 423, 453, 468, 477, 478, 480, 485, 488, 499, 500, 505, 521, 527, 531, 567, 572, 585, 592, 604, 605, 611, 624, 636, 637, 638, 641, 646, 648, 651, 656, 662, 668, 670, 671, 673, 677, 679, 683, 690, 694, 698, 714, 728, 730, 755, 763, 764, 780, 785, 787, 788, 791, 792, 794, 809, 818, 830, 833, 838, 839, 873, 878, 882, 890, 896, 902, 905, 914, 917, 920, 921, 931, 932, 937, 939, 946, 968, 970, 979, 986, 987, 992, 999, 1005, 1009, 1011, 1025, 1028, 1040, 1041, 1042, 1043, 1048, 1064, 1069, 1078, 1082, 1084, 1109, 1112, 1118, 1123, 1126, 1128, 1132, 1144, 1170, 1173, 1176, 1181, 1183, 1209, 1210, 1214, 1230, 1233, 1234, 1235, 1236, 1261, 1263, 1275, 1279, 1283, 1292, 1297, 1442, 601, 654, 1122, 1198, 332, 777, 666, 301, 608, 1081, 426, 924, 577, 265, 1047, 1187, 219, 959, 901, 552, 888, 1140, 64, 935, 1008, 1213, 1015, 1016, 1017, 1018, 374, 524, 1153, 1192, 1194, 841, 964, 996, 233, 1211, 721, 1190, 1188, 806, 907, 50, 418, 71, 400, 1168, 1116, 1093, 53, 370, 596, 45, 712, 1148, 323, 825, 402, 713, 195, 176, 508, 123, 1197, 91, 602, 245, 535, 230, 231, 232, 991, 248, 92, 187, 502, 1124, 537, 1199, 941, 553, 302, 289, 173, 134, 1137, 948, 509, 619, 1034, 447, 707, 708, 1285, 928, 1113, 802, 803, 192, 613, 614, 393, 506, 369, 1206, 264, 699, 759, 722, 1100, 350, 748, 1220, 657, 658, 659, 854, 565, 734, 735, 736, 737, 228, 642, 1054, 490, 391, 425, 513, 898, 424, 105, 993, 732, 580, 1256, 704, 695, 696, 238, 706, 1253, 1327, 1032, 319, 1252, 469, 541, 731, 1019, 1023, 1215, 247, 1191, 856, 51, 971, 603, 1003, 270, 267, 268, 1242, 47, 1089, 831, 530, 1091, 193, 578, 322, 1161, 126, 197, 884, 146, 62, 1095, 273, 29, 63, 412, 346, 622, 371, 740, 715, 635, 1134, 1135, 520, 936, 990, 1002, 911, 49, 978, 586, 1004, 1036, 653, 1086, 1180, 776, 288, 1245, 72, 625, 1221, 1087, 984, 253, 398, 1125, 1269, 773, 80, 380, 224, 222, 1010, 957, 282, 847, 1151, 1083, 1268, 711, 685, 686, 1104, 663, 664, 863, 799, 1224, 1225, 771, 631, 494, 1250, 652, 1106, 383, 52, 1278, 861, 906, 872, 382, 409, 969, 139, 1202, 28, 410, 1050, 915, 544, 299, 769, 540, 1007, 1305, 1306, 612, 1203, 1216, 620, 300, 1058, 600, 384, 90, 811, 1138, 967, 795, 1154, 655, 610, 895, 136, 944, 255, 415, 866, 954, 720, 83, 1141, 86, 88, 246, 801, 498, 1039, 574, 495, 1265, 1177, 1178, 128, 1105, 1266, 97, 703, 526, 1237, 650, 98, 427, 496, 1286, 897, 649, 401, 963, 876, 275, 800, 1012, 102, 1238, 475, 1090, 261, 739, 356, 877, 34, 35, 249, 1196, 1276, 1294, 927, 375, 162, 333, 943, 293, 903, 1111, 797, 416, 955, 1145, 514, 100, 1273, 1157, 860, 942, 961, 287, 1021, 1033, 998, 733, 154, 353, 1150, 832, 474, 471, 962, 1037, 1053, 579, 211, 404, 428, 503, 493, 1120, 837, 406, 351, 528, 559, 217, 529, 58, 566, 588, 589, 590, 591, 593, 594, 891, 949, 1146, 1267, 518, 568, 842, 744, 106, 103, 104, 385, 956, 70, 850, 250, 1097, 904, 75, 199, 328, 182, 294, 77, 751, 810, 1280, 1300, 1077, 858, 362, 1207, 874, 131, 1218, 1130, 972, 710, 934, 587, 645, 855, 178, 790, 466, 94, 276, 952, 548, 115, 997, 116, 753, 606, 320, 114, 181, 1182, 1065, 562, 180, 789, 1067, 1129, 975, 626, 89, 523, 243, 223, 1617, 186, 555, 598, 1119, 702, 321, 338, 137, 1051, 660, 1088, 741, 742, 1029, 1136, 95, 844, 973, 1162, 1000, 394, 142, 561, 277, 269, 304, 558, 132, 804, 1044, 822, 852, 476, 862, 96, 396, 254, 373, 875, 312, 313, 900, 220, 563, 330, 367, 473, 515, 909, 112, 121, 762, 564, 1074, 1075, 20, 124, 357, 807, 834, 1066, 184, 697, 857, 634, 980, 782, 783, 125, 1096, 225, 812, 813, 1055, 913, 1185, 368, 349, 570, 781, 354, 386, 387, 525, 66, 388, 235, 236, 549, 251, 278, 868, 1006, 1060, 516, 517, 774, 60, 166, 501, 845, 846, 597, 381, 492, 908, 607, 507, 923, 775, 403, 719, 687, 78, 1030, 1031, 185, 1147, 689). 2. Model Number: 722011. No UDI. Serial Numbers: 157, 83, 308, 95, 107, 34, 59, 23, 319, 49, 170, 137, 173, 101, 84, 155, 133, 234, 345, 124, 332, 102, 270, 177, 132, 325, 15, 76, 79, 17, 87, 237, 255, 50, 171, 168, 22, 275, 344, 324, 140, 163, 172, 103, 82, 154, 215, 40, 69, 272, 286, 81, 224, 184, 266, 175, 146, 164, 334, 338, 165, 80, 231, 323, 327, 179, 145, 117, 6, 355, 151, 202, 21, 203, 214, 340, 18, 301, 302, 303, 304, 305, 306, 307, 257, 265, 127, 180, 259, 181, 211, 206, 217, 314, 98, 201, 341, 318, 347, 261, 178, 183, 271, 253, 230, 317, 198, 321, 204, 94, 104, 52, 290, 291, 244, 333, 287, 39, 282, 78, 216, 158, 329, 246, 295, 56, 337, 292, 125, 225, 300, 88, 309, 86, 222, 71, 156, 233, 31, 188, 77, 99, 238, 223, 51, 153, 68, 129, 186, 118, 119, 41, 55, 263, 278, 258, 32, 89, 108, 310, 311, 139, 141, 66, 274, 296, 243, 191, 227, 273, 161, 220, 242, 166, 189, 185, 100, 190, 299, 280, 199, 194, 74, 75, 228, 250, 251, 91, 213, 12, 212, 113, 152, 112, 111, 115, 114, 116, 43, 277, 313, 26, 27, 29. 3. Model Number: 722012. UDI (Serial Numbers): (01)00884838059054(21)1356 (1356); (01)00884838059054(21)191 (191); (01)00884838059054(21)351 (351); (01)00884838059054(21)583 (583); (01)00884838059054(21)655 (655); (01)00884838059054(21)796 (796); (01)00884838059054(21)950 (950); No UDI (1117, 1001, 1673, 1712, 585, 517, 1545, 70, 1715, 504, 757, 935, 1250, 1613, 2006, 1733, 1734, 2201, 1922, 741, 92, 1034, 1359, 1848, 865, 1908, 1166, 1113, 1659, 494, 335, 1804, 1473, 1408, 1401, 1588, 576, 595, 590, 2302, 1279, 2151, 481, 348, 543, 893, 1187, 1455, 2085, 2202, 423, 525, 1798, 1656, 399, 990, 1871, 951, 1652, 894, 1590, 1597, 596, 1054, 1189, 1099, 823, 1037, 566, 72, 448, 1606, 924, 1610, 686, 1108, 1294, 568, 500, 459, 469, 1422, 1077, 1537, 1035, 285, 243, 84, 1317, 1649, 1895, 1900, 1723, 878, 919, 1581, 2221, 2222, 1515, 1803, 1731, 1017, 1020, 380, 1811, 2171, 1497, 782, 418, 1053, 1558, 1336, 601, 1014, 684, 1265, 331, 389, 732, 751, 1324, 2276, 2086, 1285, 2219, 298, 814, 815, 1120, 278, 297, 1713, 1506, 580, 1098, 1852, 1307, 1858, 377, 758, 1598, 172, 2054, 856, 1690, 1407, 480, 690, 579, 1327, 733, 1091, 1254, 797, 213, 735, 1280, 2138, 828, 458, 1118, 740, 1249, 542, 1033, 2265, 490, 1075, 2203, 1321, 1396, 605, 606, 444, 66, 2092, 1269, 67, 244, 1002, 1022, 1023, 2264, 2234, 1805, 1192, 1823, 483, 484, 615, 2150, 892, 1478, 1262, 1688, 1961, 1732, 1062, 649, 650, 767, 1104, 1641, 1277, 1975, 276, 316, 1335, 1461, 159, 489, 766, 639, 669, 671, 1583, 451, 625, 651, 1047, 107, 1447, 99, 670, 124, 1616, 1660, 1083, 1807, 1924, 511, 1714, 122, 101, 75, 1406, 352, 492, 472, 1105, 642, 643, 1270, 577, 578, 1306, 855, 1456, 2269, 71, 1078, 2285, 1793, 1797, 1216, 948, 387, 498, 1662, 104, 1868, 1743, 369, 1239, 1561, 464, 2293, 1960, 616, 1508, 1084, 1038, 1058, 318, 1990, 1796, 1026, 1337, 685, 656, 1560, 1441, 2926, 1854, 2288, 473, 2068, 624, 1586, 1617, 1904, 1354, 1918, 1906, 160, 383, 560, 68, 114, 1948, 1195, 518, 888, 1329, 1992, 323, 379, 792, 1003, 1219, 89, 115, 2003, 2004, 135, 1832, 2133, 2166, 562, 1941, 760, 859, 1593, 1608, 1947, 1950, 731, 520, 1328, 848, 2292, 1946, 274, 1063, 611, 1587, 1226, 613, 1271, 1842, 205, 2147, 1799, 1665, 187, 1513, 2015, 2207, 319, 2161, 1116, 1993, 208, 943, 150, 1281, 1945, 1919, 110, 337, 1735, 1326, 788, 488, 2002, 132, 1218, 1519, 1533, 336, 694, 1334, 2013, 1405, 1141, 923, 925, 533, 406, 1293, 513, 1446, 1625, 1675, 1683, 1148, 1181, 207, 1305, 790, 499, 1716, 934, 1726, 85, 476, 1657, 1045, 1531, 1630, 851, 1827, 277, 1923, 1516, 404, 453, 321, 334, 197, 1869, 2215, 574, 1611, 1559, 2924, 1686, 1635, 858, 1224, 1345, 86, 2057, 1430, 1255, 1599, 302, 610, 1209, 920, 666, 1989, 2014, 2022, 1126, 1943, 1188, 1073, 1404, 1664, 103, 1272, 2063, 1467, 1596, 326, 926, 675, 756, 1027, 424, 1894, 2145, 2146, 2148, 2065, 1527, 1477, 2227, 1177, 1645, 93, 95.1167, 1595, 88, 1672, 909, 1210, 1512, 1101, 787, 1399, 123, 126, 136, 952, 953, 954, 1005, 1661, 2156, 320, 186, 455, 570, 1622, 1115, 1316, 1212, 975, 976, 1032, 1746, 986, 971, 1594, 233, 239, 640, 446, 69, 430, 1542, 76, 1607, 1211, 1004, 2046, 974, 1044, 505, 310, 192, 314, 852, 224, 1016, 1097, 315, 1532, 1964, 2135, 1357, 589, 1092, 248, 846, 963, 964, 1658, 1344, 1902, 328, 781, 1909, 1132, 1361, 1504, 1433, 229, 142, 962, 1260, 1094, 837, 333, 1681, 1815, 922, 1333, 774, 1492, 2047, 1655, 1855, 2098, 930, 1052, 1421, 627, 1428, 1994, 1802, 1956, 931, 1817, 309, 2168, 1168, 1562, 967, 167, 910, 632, 863, 957, 289, 137, 1040, 1444, 409, 993, 995, 57, 1072, 516, 2042, 1332, 809, 795, 1460, 170, 914, 1156, 347, 1291, 1971, 664, 180, 1376, 2107, 139, 350, 628, 1865, 550, 1872, 759, 1464, 1363, 1457, 2020, 553, 288, 1980, 1580, 1708, 1495, 1244, 556, 546, 806, 1289, 1284, 1144, 1145, 223, 1081, 1436, 726, 755, 648, 1391, 1972, 215, 871, 629, 1631, 125, 432, 966, 1157, 1623, 1702, 916, 2112, 2123, 2118, 98, 1297, 2103, 211, 1147, 1303, 2300, 2120, 199, 1833, 1874, 631, 1887, 891, 1382, 1471, 1818, 2270, 1878, 1462, 1384, 798, 165, 210, 1203, 1966, 1178, 1999, 793, 794, 667, 2196, 1669, 502, 1021, 1142, 1879, 1703, 2031, 659, 1572, 1076, 2291, 1093, 479, 1043, 1626, 2056, 608, 1624, 216, 2204, 1138, 839, 805, 381, 752, 555, 1548, 1814, 1808, 979, 390, 261, 269, 2188, 1056, 866, 1699, 619, 1267, 311, 477, 1174, 1485, 59, 541, 1415, 764, 388, 876, 1150, 1701, 1693, 2080, 1986, 772, 1241, 1819, 1898, 911, 978, 2093, 116, 56, 2181, 599, 2034, 1071, 622, 941, 1088, 1927, 1412, 181, 548, 582, 588, 1567, 1367, 1143, 1423, 1268, 279, 1592, 1122, 2028, 1375, 1059, 400, 166, 746, 771, 770, 2177, 280, 1202, 753, 2274, 2198, 1028, 581, 1309, 810, 1086, 524, 1390, 1721, 374, 738, 360, 992, 219, 1825, 1839, 652, 2102, 2127, 1834, 829, 1434, 665, 742, 737, 1360, 1349, 391, 415, 2095, 730, 1967, 1996, 763, 332, 867, 902, 917, 1366, 2096, 1201, 587, 1061, 2072, 1691, 1861, 407, 1741, 833, 554, 214, 779, 799, 2101, 1228, 2117, 534, 1348, 813, 1740, 1530, 998, 1266, 680, 1997, 441, 60, 803, 2195, 895, 509, 1736, 355, 1381, 112, 478, 227, 785, 291, 185, 1450, 540, 1976, 2016, 1591, 443, 106, 127, 382, 661, 727, 1137, 1795, 1128, 1377, 392, 1134, 1130, 296, 266, 1074, 376, 2067, 152, 1978, 663, 1678, 569, 1481, 2349, 436, 1929, 1654, 287, 944, 2176, 1413, 188, 2140, 1998, 635, 1089, 1668, 1256, 2272, 660, 1671, 743, 58, 128, 1287, 1079, 61, 982, 693, 1388, 184, 1479, 1801, 2074, 1342, 405, 1535, 1882, 1127, 980, 2058, 1705, 260, 539, 2278, 748, 1435, 528, 506, 1885, 1926, 2073, 572, 688, 903, 410, 901, 869, 728, 91, 218, 419, 275, 830, 2094, 1411, 1222, 1397, 272, 2153, 1711, 1965, 1480, 1937, 1298, 414, 338, 1237, 1151, 1152, 1347, 1977, 1836, 1131, 1242, 573, 1484, 117, 2082, 1695, 817, 1536, 630, 593, 773, 1214, 936, 679, 1816, 544, 2392, 1862, 1364, 2033, 1296, 2200, 2017, 2208, 1175, 1233, 1653, 1463, 955, 2087, 1698, 1064, 1958, 271, 1135, 1385, 870, 507, 2182, 1223, 1288, 2008, 398, 1234, 965, 662, 868, 1483, 1489, 647, 1232, 2175, 1984, 222, 840, 175, 176, 1261, 437, 557, 1082, 1633, 426, 672, 1379, 872, 1973, 1822, 144, 692, 1820, 1821, 1312, 2077, 161, 140, 1696, 678, 780, 73, 74, 1292, 1466, 34, 119, 2043, 155, 1837, 118, 1451, 598, 1158, 1153, 1663, 1546, 835, 177, 1720, 749, 1400, 1897, 1246, 1245, 1939, 897, 2137, 111, 2206, 1585, 880, 1845, 777, 403, 26, 523, 1951, 2144, 1494, 2921, 385, 1930, 609, 1748, 1425, 1988, 2000, 148, 592, 1870, 674, 1049, 510, 2294, 2316, 1829, 1838, 1915, 1172, 687, 1710, 80, 1987, 428, 602, 623, 2220, 775, 988, 1582, 1979, 928, 807, 1322, 529, 1891, 465, 1719, 808, 1048, 1024, 824, 745, 617, 985, 1259, 1136, 2927, 1534, 1350, 1452, 1620, 802, 913, 1180, 354, 1813, 522, 983, 1121, 421, 1867, 1198, 416, 1840, 1812, 2066, 212, 933, 1676, 1634, 2059, 1541, 1338, 1520, 1955, 178, 249, 1418, 1251, 1176, 1417, 1899, 1454, 1055, 1962, 1015, 1124, 1547, 2266, 1230, 1543, 1609, 1744, 2110, 1553, 2922, 940, 1639.657, 2141, 1619, 143, 1368, 1149, 292, 887, 113, 105, 1742, 2280, 1949, 475, 1850, 1901, 845, 2100, 273, 1319, 2036, 468, 1341, 545, 526, 769, 1554, 1651, 932, 1060, 1647, 1220, 1200, 1932, 1931, 1935, 491, 514, 294, 1227, 236, 408, 1847, 2089, 873, 874, 1046, 994, 1745, 173, 1700, 2097, 341, 342, 1648, 853, 744, 786, 1355, 1925, 1313, 1991, 373, 1171, 1551, 1794, 2091, 1615, 1439, 1911, 149, 1080, 1, 618, 471, 1844, 1697, 972, 353, 2114, 1283, 1448, 267, 1123, 1206, 497, 425, 1330, 1888, 1050, 2253, 1273, 1974, 1907, 818, 621, 890, 1374, 1747, 162, 1371, 1125, 1031, 1085, 1864, 1565, 179, 946, 1372, 945, 1449, 1517, 1186, 2078, 1318, 1264, 1351, 1640, 1667, 1248, 1637, 1576, 1575, 1493, 646, 1577, 1579, 1574, 329, 2069, 1914, 1173, 343, 304, 306, 307, 308, 1684, 433, 413, 1159, 429, 2010, 1490, 1670, 921, 882, 1051, 2347, 567, 422, 1304, 849, 1853, 1057, 1295, 2055, 603, 1090, 463, 1841, 1129, 1959, 604, 850, 454, 460, 2053, 812, 1468, 1921, 1896, 1982, 1503, 295, 1969, 2130, 411, 183, 375, 969, 1628, 634, 1555, 847, 1666, 240, 1320, 1706, 1528, 241, 242, 157, 156, 804, 1432, 1008, 1068, 658). 4. Model Number: 722013. No UDI. Serial Numbers: 505, 217, 313, 492, 34, 153, 410, 360, 290, 74, 83, 580, 160, 404, 126, 244, 106, 346, 461, 307, 1, 442, 435, 293, 297, 46, 289, 638, 342, 69, 267, 214, 402, 113, 86, 603, 82, 149, 424, 140, 254, 255, 266, 475, 518, 519, 303, 194, 115, 150, 496, 599, 441, 491, 124, 151, 179, 550, 377, 575, 562, 178, 309, 601, 327, 73, 77, 264, 78, 308, 141, 357, 85, 40, 216, 637, 422, 174, 392, 109, 218, 228, 511, 389, 152, 479, 262, 61, 401, 497, 84, 125, 241, 564, 573, 426, 556, 42, 100, 495, 500, 96, 515, 105, 428, 427, 531, 375, 316, 278, 347, 205, 246, 43, 639, 35, 480, 304, 147, 15, 52, 283, 168, 171, 265, 272, 332, 509, 444, 400, 292, 117, 598, 415, 454, 222, 215, 335, 416, 47, 384, 559, 361, 358, 549, 476, 133, 36, 315, 31, 490, 417, 48, 276, 247, 618, 390, 157, 53, 238, 525, 17, 528, 224, 76, 585, 271, 172, 50, 544, 570, 359, 296, 530, 232, 478, 457, 547, 183, 184, 185, 186, 187, 188, 189, 190, 413, 437, 107, 138, 227, 318, 341, 225, 29, 508, 103, 434, 473, 135, 551, 557, 541, 294, 128, 388, 207, 258, 429, 95, 411, 353, 542, 533, 523, 182, 574, 139, 137, 250, 563, 502, 450, 436, 275, 524, 354, 510, 395, 458, 261, 595, 586, 535, 569, 425, 334, 344, 355, 356, 465, 345, 501, 522, 64, 396, 208, 408, 555, 394, 57, 520, 493, 201, 366, 527, 65, 449, 167, 70, 312, 321, 161, 324, 311, 176, 181, 554, 104, 459, 130, 325, 470, 440, 144, 28, 385, 403, 81, 134, 487, 256, 374, 405, 380, 367, 108, 352, 71, 455, 468, 60, 301, 299, 191, 300, 302, 239, 243, 110, 111, 158, 378, 223, 343, 165, 453, 291, 532, 364, 370, 22, 193, 336, 132, 236, 163, 2266, 368, 177, 98, 156, 443, 259, 540. 5. Model Number: 722022. No UDI. Serial Number: 6. 6. Model Number: 722023. No UDI. Serial Numbers: 27, 116, 12, 175, 98, 35, 72, 142, 46, 64, 110, 49, 124, 66, 117, 109, 42, 138, 82, 83, 127, 75, 118, 41, 33, 146, 121, 119, 70, 112, 47, 137, 25, 114, 105, 90, 107, 74, 5, 9, 4, 14, 13, 16, 132, 88, 73, 99, 38, 141, 32, 84, 68, 134, 79, 52, 56, 69, 37, 21, 53, 123, 157, 102, 126, 145, 144, 15, 135, 152, 149, 60, 150, 108, 100, 95, 34, 54, 85, 87, 94, 57, 30, 130, 58, 86, 125, 96, 3, 103, 10, 31, 106, 67. 7. Model Number: 722025. No UDI. Serial Numbers: 3, 2, 6.
Why was this product recalled?
Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?
Philips Medical Systems notified consignees on 11/11/2024 via letter. Consignees were instructed that affected systems may continue to be used in accordance with their Indications for Use and Instructions for Use (IFU), to keep the Urgent Medical Device Correction letter with the documentation of the system until Philips corrects the system, circulate this notice to all users of the system so that they are aware of the issue, and if you experience erratic movement of the C-arm and/or if you see a portion of black area on the X-ray image, please call your local Philips representative to report the event. Philips will contact all affected customers to arrange for a Field Service Engineer visit to install an updated Lifetime Extension Kit. Consignees were also instructed to notify customers if further distributed and inform Philips about the transfer through the email account: [email protected]. Philips requested consignees complete and return the provided response form. On 11/24/2025, Philips sent a follow up letter to consignees to provide an update to the Urgent Medical Device Correction (MDC) issued on 11/11/2024. The updated letter informed consignees that after further technical investigation on both the hydrogen embrittlement and fatigue conditions, Philips has confirmed that these failure mechanisms do not apply to the bolts in question. Therefore, replacement is not required, even after 10 years of use. As a result, the risks and actions outlined in the original MDC no longer apply to affected systems.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0676-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology