Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory electrocardiogram. Recall
Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory electrocardiogram.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory electrocardiogram.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0684-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 25, 2024
- FDA report date
- Dec 25, 2024
- Quantity in commerce
- 13,202 devices
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory electrocardiogram.
Affected model & UDI checker
Search the exact identifiers published for Z-0684-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 98700 | Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory electrocardiogram. | Check source notice | Not specified |
| Not specified | Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory electrocardiogram. | Check source notice | 10841522128851 |
FDA code information
Model Number: 98700 UDI-DI code: 10841522128851 Serial Numbers: All serial numbers manufactured between April 2021 to November 2024. Serial Numbers pending.
Why was this product recalled?
Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Spacelabs Healthcare, Ltd. take?
On 11/25/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter to customers inform them that the Eclipse Pro and Eclipse MINI cardiac Holter monitors may have mismatched serial numbers between the label and what is programmed in the devices. This is due to a manufacturing error. Customers are instructed to: complete the following steps for each Eclipse device under your control. Following these steps will allow you to check if your device is affected. 1. Visit https://eclipseutils.cardiology.spacelabs.com/ 2. The first time you use this site you will need to register for an account. Click Register and create a username and password. 3. Plug Eclipse (Mini or Pro) device into the computer via USB. 4. Select Serial Check in the menu and click CONNECT NEW RECORDER 5. Select the device from the list and click Connect 6. Click VERIFY SERIAL NUMBER once the device is connected 7. When prompted, enter the serial number printed on the label of the device and click CONTINUE 8. If the serial number you entered on the website does not match the serial number loaded into the device, you will see the warning. Confirm that you do not have a typo in the serial number you entered on the website. If you have a typo, click GO BACK . If correctly entered the serial number click CONFIRM . 9. If the serial number entered on the website matches the serial number loaded into the device, you will see the confirmation below. Confirm that you do not have a typo in the serial number you entered on the website. If you have a typo, please click GO BACK . If you have correctly entered the serial number click CONFIRM . 10. Repeat the previous steps for all your eclipse devices. Questions or assistance - contact Spacelabs service at 1-800-522-7025 or visit us at https://spacelabshealthcare.com/support/ Follow-up Updated Customer Letter sent 04/04/2025
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0684-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology