Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Recall
Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: N/A
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: K-Systems Product Name: G73 Dry Bath Incubator. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0687-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 15, 2024
- FDA report date
- Dec 25, 2024
- Quantity in commerce
- 106 units
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: N/A
Affected model & UDI checker
Search the exact identifiers published for Z-0687-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| K22065 | Brand Name: K-Systems Product Name: G73 Dry Bath Incubator | Check source notice | Not specified |
FDA code information
Version or Model: K22065; Primary DI Number: 00888937025040; Serial Numbers: 2202KH46, 2202KH47, 2202KH48, 2202KH49, 2202KH50, 2202KH51, 2202KH52, 2202KH53, 2202KH54, 2202KH55, 2208KH01, 2208KH02, 2208KH03, 2208KH04, 2208KH05, 2208KH06, 2208KH07, 2208KH08, 2208KH09, 2208KH10, 2208KH11, 2208KH12, 2208KH13, 2208KH14, 2208KH15, 2208KH16, 2208KH17, 2208KH18, 2208KH19, 2208KH20, 2235KH01, 2235KH02, 2235KH03, 2235KH04, 2235KH05, 2235KH06, 2235KH07, 2235KH08, 2235KH09, 2235KH10, 2235KH11, 2235KH12, 2235KH13, 2235KH14, 2235KH15, 2235KH16, 2235KH17, 2235KH18, 2235KH19, 2235KH20, 2239KH11, 2239KH12, 2239KH13, 2239KH14, 2239KH15, 2239KH16, 2239KH17, 2239KH18, 2239KH19, 2239KH20, 2242KH11, 2242KH12, 2242KH13, 2242KH14, 2242KH15, 2242KH16, 2242KH17, 2242KH18, 2242KH19, 2242KH20, 2242KH21, 2242KH22, 2242KH23, 2242KH24, 2242KH25, 2246KH01, 2246KH02, 2246KH03, 2246KH04, 2246KH05, 2246KH06, 2246KH07, 2246KH08, 2246KH09, 2246KH10, 2246KH11, 2246KH12, 2246KH13, 2246KH14, 2246KH15, 2246KH16, 2246KH17, 2246KH18, 2246KH19, 2246KH20, 2246KH21, 2246KH22, 2246KH23, 2246KH24, 2246KH25, 2246KH26, 2246KH27, 2246KH28, 2246KH29, 2246KH30, 2246KH31, 2246KH32, 2246KH33, 2246KH34, 2246KH35, 2246KH36, 2246KH37, 2246KH38, 2246KH39, 2246KH40, 2246KH41, 2246KH42, 2246KH43, 2246KH44, 2246KH45, 2246KH46, 2246KH47, 2246KH48, 2246KH49, 2246KH50, 2246KH51, 2246KH52, 2246KH53, 2246KH54, 2246KH55, 2303KH01, 2303KH02, 2303KH03, 2303KH04, 2303KH05, 2303KH06, 2303KH07, 2303KH08, 2303KH09, 2303KH10, 2303KH11, 2303KH12, 2303KH13, 2303KH14, 2303KH15, 2303KH16, 2303KH17, 2303KH18, 2303KH19, 2303KH20, 2303KH21, 2303KH22, 2303KH23, 2303KH24, 2303KH25, 2303KH26, 2303KH27, 2303KH28, 2303KH29, 2303KH30, 2303KH31, 2303KH32, 2303KH33, 2303KH34, 2303KH35, 2303KH36, 2303KH37, 2303KH38, 2303KH39, 2303KH40, 2311KH01, 2311KH02, 2311KH03, 2311KH04, 2311KH05, 2311KH06, 2311KH07, 2311KH08, 2311KH09, 2311KH10, 2311KH11, 2311KH12, 2311KH13, 2311KH14, 2311KH15, 2313KH11, 2313KH12, 2313KH13, 2313KH14, 2313KH15, 2313KH16, 2313KH17, 2313KH18, 2313KH19, 2313KH20, 2313KH21, 2313KH22, 2313KH23, 2313KH24, 2313KH25, 2316KH01, 2316KH02, 2316KH03, 2316KH04, 2316KH05, 2316KH06, 2316KH07, 2316KH08, 2316KH09, 2316KH10, 2316KH11, 2316KH12, 2316KH13, 2316KH14, 2316KH15, 2322KH11, 2322KH12, 2322KH13, 2322KH14, 2322KH15, 2322KH16, 2322KH17, 2322KH18, 2322KH19, 2322KH20, 2322KH21, 2322KH22, 2322KH23, 2322KH24, 2322KH25, 2326KH01, 2326KH02, 2326KH03, 2326KH04, 2326KH05, 2326KH06, 2326KH07, 2326KH08, 2326KH09, 2326KH10, 2326KH11, 2326KH12, 2326KH13, 2326KH14, 2326KH15, 2326KH16, 2326KH17, 2326KH18, 2326KH19, 2326KH20, 2327KH30, 2327KH31, 2327KH32, 2327KH33, 2327KH34, 2327KH35, 2327KH36, 2327KH37, 2327KH38, 2327KH39, 2327KH40, 2327KH41, 2327KH42, 2327KH43, 2327KH44, 2332KH11, 2332KH12, 2332KH13, 2332KH14, 2332KH15, 2332KH16, 2332KH17, 2332KH18, 2332KH19, 2332KH20.
Why was this product recalled?
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
What actions did CooperSurgical, Inc. take?
On November 15, 2024 URGENT MEDICAL DEVICE RECALL emails were sent to customers. On November 20, 2024 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions that should be taken by the customer/user/distributor in order to prevent risks for patients or users: G85 Mini Incubators with serial numbers listed in the Appendix are impacted and require repair. 1. Inspect your inventory, identify, and discontinue use. 2. Complete the attached Urgent Medical Device Recall Response Form and return the form to CooperSurgical as indicated at the top of the form. Note: Please complete and return the form even if you do not have any affected product in your inventory so that we may document receipt of this Urgent Medical Device Recall. 3. Contact your local Service Representative to arrange a repair at no cost to you. T45, T47, R65, and G73 with serial numbers listed in the Appendix may be impacted and require inspection to determine the need for repair. 1. Inspect your inventory, identify, and discontinue use. 2. Complete the attached Urgent Medical Device Recall Response Form and return the form to CooperSurgical as indicated at the top of the form. Note: Please complete and return the form even if you do not have any affected product in your inventory so that we may document receipt of this Urgent Medical Device Recall. 3. Contact your local Service Representative to arrange an inspection of the affected product. If your product is affected, the product will be repaired at no cost to you. 6. Actions planned by CooperSurgical to correct the problem CooperSurgical is currently investigating this issue and determined a field action notice to customers is needed to inspect and repair all affected products in the field. CooperSurgical is committed to high-quality, safe, and effective products. This corrective action has been initiated to ensure this failure mode does not reoccur and future potential patient harm can be avoided. If you need any further
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0687-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology