FDA RECALL RECORD · Z-0687-2025

Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Recall

Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: K-Systems Product Name: G73 Dry Bath Incubator. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0687-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 15, 2024
FDA report date
Dec 25, 2024
Quantity in commerce
106 units
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-0687-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
K22065Brand Name: K-Systems Product Name: G73 Dry Bath IncubatorCheck source noticeNot specified
FDA code information

Version or Model: K22065; Primary DI Number: 00888937025040; Serial Numbers: 2202KH46, 2202KH47, 2202KH48, 2202KH49, 2202KH50, 2202KH51, 2202KH52, 2202KH53, 2202KH54, 2202KH55, 2208KH01, 2208KH02, 2208KH03, 2208KH04, 2208KH05, 2208KH06, 2208KH07, 2208KH08, 2208KH09, 2208KH10, 2208KH11, 2208KH12, 2208KH13, 2208KH14, 2208KH15, 2208KH16, 2208KH17, 2208KH18, 2208KH19, 2208KH20, 2235KH01, 2235KH02, 2235KH03, 2235KH04, 2235KH05, 2235KH06, 2235KH07, 2235KH08, 2235KH09, 2235KH10, 2235KH11, 2235KH12, 2235KH13, 2235KH14, 2235KH15, 2235KH16, 2235KH17, 2235KH18, 2235KH19, 2235KH20, 2239KH11, 2239KH12, 2239KH13, 2239KH14, 2239KH15, 2239KH16, 2239KH17, 2239KH18, 2239KH19, 2239KH20, 2242KH11, 2242KH12, 2242KH13, 2242KH14, 2242KH15, 2242KH16, 2242KH17, 2242KH18, 2242KH19, 2242KH20, 2242KH21, 2242KH22, 2242KH23, 2242KH24, 2242KH25, 2246KH01, 2246KH02, 2246KH03, 2246KH04, 2246KH05, 2246KH06, 2246KH07, 2246KH08, 2246KH09, 2246KH10, 2246KH11, 2246KH12, 2246KH13, 2246KH14, 2246KH15, 2246KH16, 2246KH17, 2246KH18, 2246KH19, 2246KH20, 2246KH21, 2246KH22, 2246KH23, 2246KH24, 2246KH25, 2246KH26, 2246KH27, 2246KH28, 2246KH29, 2246KH30, 2246KH31, 2246KH32, 2246KH33, 2246KH34, 2246KH35, 2246KH36, 2246KH37, 2246KH38, 2246KH39, 2246KH40, 2246KH41, 2246KH42, 2246KH43, 2246KH44, 2246KH45, 2246KH46, 2246KH47, 2246KH48, 2246KH49, 2246KH50, 2246KH51, 2246KH52, 2246KH53, 2246KH54, 2246KH55, 2303KH01, 2303KH02, 2303KH03, 2303KH04, 2303KH05, 2303KH06, 2303KH07, 2303KH08, 2303KH09, 2303KH10, 2303KH11, 2303KH12, 2303KH13, 2303KH14, 2303KH15, 2303KH16, 2303KH17, 2303KH18, 2303KH19, 2303KH20, 2303KH21, 2303KH22, 2303KH23, 2303KH24, 2303KH25, 2303KH26, 2303KH27, 2303KH28, 2303KH29, 2303KH30, 2303KH31, 2303KH32, 2303KH33, 2303KH34, 2303KH35, 2303KH36, 2303KH37, 2303KH38, 2303KH39, 2303KH40, 2311KH01, 2311KH02, 2311KH03, 2311KH04, 2311KH05, 2311KH06, 2311KH07, 2311KH08, 2311KH09, 2311KH10, 2311KH11, 2311KH12, 2311KH13, 2311KH14, 2311KH15, 2313KH11, 2313KH12, 2313KH13, 2313KH14, 2313KH15, 2313KH16, 2313KH17, 2313KH18, 2313KH19, 2313KH20, 2313KH21, 2313KH22, 2313KH23, 2313KH24, 2313KH25, 2316KH01, 2316KH02, 2316KH03, 2316KH04, 2316KH05, 2316KH06, 2316KH07, 2316KH08, 2316KH09, 2316KH10, 2316KH11, 2316KH12, 2316KH13, 2316KH14, 2316KH15, 2322KH11, 2322KH12, 2322KH13, 2322KH14, 2322KH15, 2322KH16, 2322KH17, 2322KH18, 2322KH19, 2322KH20, 2322KH21, 2322KH22, 2322KH23, 2322KH24, 2322KH25, 2326KH01, 2326KH02, 2326KH03, 2326KH04, 2326KH05, 2326KH06, 2326KH07, 2326KH08, 2326KH09, 2326KH10, 2326KH11, 2326KH12, 2326KH13, 2326KH14, 2326KH15, 2326KH16, 2326KH17, 2326KH18, 2326KH19, 2326KH20, 2327KH30, 2327KH31, 2327KH32, 2327KH33, 2327KH34, 2327KH35, 2327KH36, 2327KH37, 2327KH38, 2327KH39, 2327KH40, 2327KH41, 2327KH42, 2327KH43, 2327KH44, 2332KH11, 2332KH12, 2332KH13, 2332KH14, 2332KH15, 2332KH16, 2332KH17, 2332KH18, 2332KH19, 2332KH20.

View the original FDA recall notice

Why was this product recalled?

The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.

What actions did CooperSurgical, Inc. take?

On November 15, 2024 URGENT MEDICAL DEVICE RECALL emails were sent to customers. On November 20, 2024 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions that should be taken by the customer/user/distributor in order to prevent risks for patients or users: G85 Mini Incubators with serial numbers listed in the Appendix are impacted and require repair. 1. Inspect your inventory, identify, and discontinue use. 2. Complete the attached Urgent Medical Device Recall Response Form and return the form to CooperSurgical as indicated at the top of the form. Note: Please complete and return the form even if you do not have any affected product in your inventory so that we may document receipt of this Urgent Medical Device Recall. 3. Contact your local Service Representative to arrange a repair at no cost to you. T45, T47, R65, and G73 with serial numbers listed in the Appendix may be impacted and require inspection to determine the need for repair. 1. Inspect your inventory, identify, and discontinue use. 2. Complete the attached Urgent Medical Device Recall Response Form and return the form to CooperSurgical as indicated at the top of the form. Note: Please complete and return the form even if you do not have any affected product in your inventory so that we may document receipt of this Urgent Medical Device Recall. 3. Contact your local Service Representative to arrange an inspection of the affected product. If your product is affected, the product will be repaired at no cost to you. 6. Actions planned by CooperSurgical to correct the problem CooperSurgical is currently investigating this issue and determined a field action notice to customers is needed to inspect and repair all affected products in the field. CooperSurgical is committed to high-quality, safe, and effective products. This corrective action has been initiated to ensure this failure mode does not reoccur and future potential patient harm can be avoided. If you need any further

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.