FDA RECALL RECORD · Z-0703-2025

OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures. Recall

OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.

FDA recall Class II

Ongoing

FDA lists this Class II recall for OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0703-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 21, 2024
FDA report date
Dec 25, 2024
Quantity in commerce
84 systems
Distribution
See key details
FDA cause
Process control

What is being recalled?

OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.

Affected model & UDI checker

Search the exact identifiers published for Z-0703-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedOEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.Check source notice00195278096944
FDA code information

Product Name: OEC 3D GTIN code: 00195278096944 Lot/Serial Numbers / UDI codes: FBMHDX00436 (01)00195278096944(11)240800(21)FBMHDX00436 FBMHDX00352 (01)00195278096944(11)240500(21)FBMHDX00352 FBMHDX00455 (01)00195278096944(11)240900(21)FBMHDX00455 FBMHDX00384 (01)00195278096944(11)240600(21)FBMHDX00384 FBMHDX00434 (01)00195278096944(11)240900(21)FBMHDX00434 FBMHDX00443 (01)00195278096944(11)240900(21)FBMHDX00443 FBMHDX00407 (01)00195278096944(11)240700(21)FBMHDX00407 FBMHDX00363 (01)00195278096944(11)240500(21)FBMHDX00363 FBMHDX00424 (01)00195278096944(11)240800(21)FBMHDX00424 FBMHDX00442 (01)00195278096944(11)240900(21)FBMHDX00442 FBMHDX00321 (01)00195278096944(11)240300(21)FBMHDX00321 FBMHDX00387 (01)00195278096944(11)240600(21)FBMHDX00387 FBMHDX00339 (01)00195278096944(11)240400(21)FBMHDX00339 FBMHDX00448 (01)00195278096944(11)240900(21)FBMHDX00448 FBMHDX00383 (01)00195278096944(11)240600(21)FBMHDX00383 FBMHDX00437 (01)00195278096944(11)240900(21)FBMHDX00437 FBMHDX00362 (01)00195278096944(11)240500(21)FBMHDX00362 FBMHDX00394 (01)00195278096944(11)240600(21)FBMHDX00394 FBMHDX00397 (01)00195278096944(11)240700(21)FBMHDX00397 FBMHDX00395 (01)00195278096944(11)240700(21)FBMHDX00395 FBMHDX00355 (01)00195278096944(11)240500(21)FBMHDX00355 FBMHDX00374 (01)00195278096944(11)240600(21)FBMHDX00374 FBMHDX00451 (01)00195278096944(11)240900(21)FBMHDX00451 FBMHDX00368 (01)00195278096944(11)240500(21)FBMHDX00368 FBMHDX00457 (01)00195278096944(11)241000(21)FBMHDX00457 FBMHDX00440 (01)00195278096944(11)240900(21)FBMHDX00440 FBMHDX00419 (01)00195278096944(11)240800(21)FBMHDX00419 FBMHDX00421 (01)00195278096944(11)240800(21)FBMHDX00421 FBMHDX00427 (01)00195278096944(11)240800(21)FBMHDX00427 FBMHDX00459 (01)00195278096944(11)240900(21)FBMHDX00459 FBMHDX00378 (01)00195278096944(11)240600(21)FBMHDX00378 FBMHDX00345 (01)00195278096944(11)240400(21)FBMHDX00345 FBMHDX00458 (01)00195278096944(11)240900(21)FBMHDX00458 FBMHDX00425 (01)00195278096944(11)240800(21)FBMHDX00425 FBMHDX00366 (01)00195278096944(11)240500(21)FBMHDX00366 FBMHDX00480 (01)00195278096944(11)241100(21)FBMHDX00480 FBMHDX00418 (01)00195278096944(11)240800(21)FBMHDX00418 FBMHDX00454 (01)00195278096944(11)240900(21)FBMHDX00454 FBMHDX00188 (01)00195278096944(11)230300(21)FBMHDX00188 FBMHDX00403 (01)00195278096944(11)240700(21)FBMHDX00403 FBMHDX00452 (01)00195278096944(11)240900(21)FBMHDX00452 FBMHDX00386 (01)00195278096944(11)240600(21)FBMHDX00386 FBMHDX00343 (01)00195278096944(11)240400(21)FBMHDX00343 FBMHDX00369 (01)00195278096944(11)240500(21)FBMHDX00369 FBMHDX00464 (01)00195278096944(11)240900(21)FBMHDX00464 FBMHDX00413 (01)00195278096944(11)240800(21)FBMHDX00413 FBMHDX00346 (01)00195278096944(11)240400(21)FBMHDX00346 FBMHDX00388 (01)00195278096944(11)240600(21)FBMHDX00388 FBMHDX00389 (01)00195278096944(11)240600(21)FBMHDX00389 FBMHDX00406 (01)00195278096944(11)240700(21)FBMHDX00406 FBMHDX00416 (01)00195278096944(11)240800(21)FBMHDX00416 FBMHDX00453 (01)00195278096944(11)240900(21)FBMHDX00453 FBMHDX00399 (01)00195278096944(11)240700(21)FBMHDX00399 FBMHDX00401 (01)00195278096944(11)240700(21)FBMHDX00401 FBMHDX00428 (01)00195278096944(11)240800(21)FBMHDX00428 FBMHDX00380 (01)00195278096944(11)240600(21)FBMHDX00380 FBMHDX00371 (01)00195278096944(11)240600(21)FBMHDX00371 FBMHDX00392 (01)00195278096944(11)240600(21)FBMHDX00392 FBMHDX00344 (01)00195278096944(11)240400(21)FBMHDX00344 FBMHDX00404 (01)00195278096944(11)240700(21)FBMHDX00404 FBMHDX00356 (01)00195278096944(11)240500(21)FBMHDX00356 FBMHDX00376 (01)00195278096944(11)240600(21)FBMHDX00376 FBMHDX00470 (01)00195278096944(11)241000(21)FBMHDX00470 FBMHDX00353 (01)00195278096944(11)240500(21)FBMHDX00353 FBMHDX00359 (01)00195278096944(11)240500(21)FBMHDX00359 FBMHDX00391 (01)00195278096944(11)240600(21)FBMHDX00391 FBMHDX00460 (01)00195278096944(11)240900(21)FBMHDX00460 FBMHDX00410 (01)00840682121729(11)240700(21)FBMHDX00410 FBMHDX00379 (01)00195278096944(11)240600(21)FBMHDX00379 FBMHDX00390 (01)00195278096944(11)240600(21)FBMHDX00390 FBMHDX00449 (01)00195278096944(11)240900(21)FBMHDX00449 FBMHDX00372 (01)00195278096944(11)240600(21)FBMHDX00372 FBMHDX00377 (01)00195278096944(11)240600(21)FBMHDX00377 FBMHDX00462 (01)00195278096944(11)240900(21)FBMHDX00462 FBMHDX00338 (01)00195278096944(11)240400(21)FBMHDX00338 FBMHDX00423 (01)00195278096944(11)240800(21)FBMHDX00423 FBMHDX00420 (01)00195278096944(11)240800(21)FBMHDX00420 FBMHDX00438 (01)00195278096944(11)240900(21)FBMHDX00438 FBMHDX00432 (01)00195278096944(11)240900(21)FBMHDX00432 FBMHDX00373 (01)00195278096944(11)240600(21)FBMHDX00373 FBMHDX00367 (01)00195278096944(11)240500(21)FBMHDX00367 FBMHDX00446 (01)00195278096944(11)240900(21)FBMHDX00446 FBMHDX00370 (01)00195278096944(11)240600(21)FBMHDX00370 FBMHDX00429 (01)00195278096944(11)240800(21)FBMHDX00429

View the original FDA recall notice

Why was this product recalled?

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE OEC Medical Systems, Inc take?

On 11/21/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that GE HealthCare has become aware that recently manufactured OEC Elite and OEC 3D mobile C-arms, as well as recently replaced X-ray tubes on certain OEC Elite, OEC 3D and OEC 9900/OEC 9800 mobile C-arms potentially have insufficient sealing that can result in an oil leak. This issue could result in loss of fluoroscopic imaging capability. Customers are instructed to: 1.Continue to use their OEC Elite, OEC 3D, OEC 9900 and/or OEC9800 mobile C-arm. 2.If oil is found leaking from the X-ray tube on their OEC Elite, OEC 3D, OEC9900 and/or OEC 9800 mobile C-arm, contact the local GE HealthCareService representative. 3.Ensure all potential users in their facility are made aware of this safety notification and the recommended actions. 4.Please complete and return the attached acknowledgement form [email protected]. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to plan and schedule for the inspection and correction if needed. for any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.