OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. Recall
OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
FDA recall Class II
Ongoing
FDA lists this Class II recall for OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0706-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 21, 2024
- FDA report date
- Dec 25, 2024
- Quantity in commerce
- 25 systems
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
Affected model & UDI checker
Search the exact identifiers published for Z-0706-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. | Check source notice | 00840682114349 |
FDA code information
Product Name: OEC 9800 GTIN code: 00840682114349 Lot/Serial Numbers / UDI codes: 82XXXX04237 (01)00840682114349(11)170600(21)82XXXX04237 8S-2249 Not Applicable-MFG date prior to UDI compliance requirement 89-1022 Not Applicable-MFG date prior to UDI compliance requirement 82-1772 Not Applicable-MFG date prior to UDI compliance requirement 82-0980 Not Applicable-MFG date prior to UDI compliance requirement 8S-1793 Not Applicable-MFG date prior to UDI compliance requirement 82-2349 Not Applicable-MFG date prior to UDI compliance requirement 82-3361 Not Applicable-MFG date prior to UDI compliance requirement 8S-1398 Not Applicable-MFG date prior to UDI compliance requirement 82-0908 Not Applicable-MFG date prior to UDI compliance requirement 82-0599 Not Applicable-MFG date prior to UDI compliance requirement 8S-0533 Not Applicable-MFG date prior to UDI compliance requirement 82-0189 Not Applicable-MFG date prior to UDI compliance requirement 8S-0186 Not Applicable-MFG date prior to UDI compliance requirement 89-2415 Not Applicable-MFG date prior to UDI compliance requirement 8S-3259 Not Applicable-MFG date prior to UDI compliance requirement 8S-3426 Not Applicable-MFG date prior to UDI compliance requirement 89-1530 Not Applicable-MFG date prior to UDI compliance requirement 82-2955 Not Applicable-MFG date prior to UDI compliance requirement 8S-0958-RC Not Applicable-MFG date prior to UDI compliance requirement 82-1154 Not Applicable-MFG date prior to UDI compliance requirement 8S-3032 Not Applicable-MFG date prior to UDI compliance requirement 8S-3145 Not Applicable-MFG date prior to UDI compliance requirement 89-1916-RC Not Applicable-MFG date prior to UDI compliance requirement 82-4134 Not Applicable-MFG date prior to UDI compliance requirement
Why was this product recalled?
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did GE OEC Medical Systems, Inc take?
On 11/21/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that GE HealthCare has become aware that recently manufactured OEC Elite and OEC 3D mobile C-arms, as well as recently replaced X-ray tubes on certain OEC Elite, OEC 3D and OEC 9900/OEC 9800 mobile C-arms potentially have insufficient sealing that can result in an oil leak. This issue could result in loss of fluoroscopic imaging capability. Customers are instructed to: 1.Continue to use their OEC Elite, OEC 3D, OEC 9900 and/or OEC9800 mobile C-arm. 2.If oil is found leaking from the X-ray tube on their OEC Elite, OEC 3D, OEC9900 and/or OEC 9800 mobile C-arm, contact the local GE HealthCareService representative. 3.Ensure all potential users in their facility are made aware of this safety notification and the recommended actions. 4.Please complete and return the attached acknowledgement form [email protected]. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to plan and schedule for the inspection and correction if needed. for any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0706-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology