Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0707-2025

Duravent Silicone Ventilation Tube. Recall

Duravent Silicone Ventilation Tube. Model Number: 240075.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Duravent Silicone Ventilation Tube.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0707-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 19, 2024
FDA report date
Dec 25, 2024
Quantity in commerce
984 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Duravent Silicone Ventilation Tube. Model Number: 240075.

Affected model & UDI checker

Search the exact identifiers published for Z-0707-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
240075.Duravent Silicone Ventilation Tube.Check source notice00521925031883
240075Duravent Silicone Ventilation Tube.Check source noticeNot specified
FDA code information

Model Number: 240075. UDI: 00521925031883 (17)331015 (10)SM405666, Lot Number: SM405666. UDI: 00521925031883 (17)331015 (10)JB362646, Lot Number: JB362646.

View the original FDA recall notice

Why was this product recalled?

Potential that Duravent Silicone Ventilation Tubes may contain incorrect product, specifically, 1.32 mm Tiny T Tubes (model: 70240076) instead of the intended Duravent Silicone Ventilation Tubes.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Olympus Corporation of the Americas take?

Olympus notified consignees on 11/19/2024 via Courier, 2-day delivery. Consignees were instructed to examine inventory for affected devices, quarantine any affected devices on hand, arrange for return of those affected units and acknowledge receipt of the customer letter through the web portal.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology