Sheath, 18 Fr., 130 mm (REF: WA2PS18S), gastroenterology-urology endoscopic access overtube. Recall

Sheath, 18 Fr., 130 mm (REF: WA2PS18S), gastroenterology-urology endoscopic access overtube.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Sheath, 18 Fr., 130 mm (REF: WA2PS18S), gastroenterology-urology endoscopic access overtube.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0715-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Nov 27, 2024
FDA report date
Jan 22, 2025
Quantity in commerce
53 units
Distribution
See key details
FDA cause
Labeling False and Misleading

What is being recalled?

Sheath, 18 Fr., 130 mm (REF: WA2PS18S), gastroenterology-urology endoscopic access overtube.

Affected model & UDI checker

Search the exact identifiers published for Z-0715-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
WA2PS18SSheath, 18 Fr., 130 mm (REF: WA2PS18S), gastroenterology-urology endoscopic access overtube.Check source noticeNot specified
Not specifiedSheath, 18 Fr., 130 mm (REF: WA2PS18S), gastroenterology-urology endoscopic access overtube.Check source notice04251303810865
FDA code information

REF: WA2PS18S; UDI-DI: 04251303810865; Lot Numbers: 159825

View the original FDA recall notice

Why was this product recalled?

If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did TROKAMED GMBH take?

On November 27, 2024, a "URGENT SAFETY FIELD NOTICE" email was sent to the distributor. Actions to be taken: - Inform the medical personnel using the product about the update of the IFU: 1. The Instructions for Use has to be exchanged. 2. Complete the enclosed acknowledgement of receipt and email to [email protected] by the provided date. - Please be informed that the assumption of risks is on your behalf when you will not send back the acknowledgement by the provided date. Disclosure of the information described herein: Please ensure in your organization that all users of the above-mentioned products and other persons to be informed are made aware of this Urgent Field Safety Notice. If you have given the products to third parties, please forward a copy of this information or inform the contact person indicated below. Please keep this information at least until the action has been completed. If you have any questions or concerns about this notification, please contact the Quality Management Representative at +49 7704 9244 60. ***Updated 2/5/2025*** On 01/14/2025 an updated "URGENT SAFETY FIELD NOTICE" letter was sent to their distributor account. What measures are to be taken by the recipient? 1. Inform the medical personnel using the product about this field safety notice. 2. Replace your current Instructions for Use for this device with the updated version provided (W9229406_01). This is not a product removal action. You may continue to use the device as per this letter and the updated Instruction for Use. 3. Acknowledge receipt according to the instructions provided by your local distributor. Please ensure in your organization that all users of the above-mentioned products and other persons to be informed are made aware of this Urgent Field Safety Notice. If you have given the products to third parties, please forward a copy of this information or inform the contact person indicated below. Please keep this information at least until the action has been

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology