Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0721-2025

Medline brand, medical procedure convenience kits, labeled as: 1) MVFF/FIBU #54-RF , SKU DYNJ21927R; 2) DIEP FLAP, SKU DYNJ910120A Recall

Medline brand, medical procedure convenience kits, labeled as: 1) MVFF/FIBU #54-RF , SKU DYNJ21927R; 2) DIEP FLAP, SKU DYNJ910120A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medline brand, medical procedure convenience kits, labeled as: 1) MVFF/FIBU #54-RF , SKU DYNJ21927R; 2) DIEP FLAP, SKU DYNJ910120A. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0721-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 6, 2024
FDA report date
Jan 1, 2025
Quantity in commerce
318 kits
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Medline brand, medical procedure convenience kits, labeled as: 1) MVFF/FIBU #54-RF , SKU DYNJ21927R; 2) DIEP FLAP, SKU DYNJ910120A

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI-DI: DYNJ21927R, 10195327104733(each), 40195327104734(case); Lot Number: 23IBF547 23GBT501 23FBU360 23EBU929 23EBL624 23EBA114 23CBN730 23ABF393 22JBK203 22IBN858 22HBO301 22GBC794 22CBI774; UDI-DI: DYNJ910120A, 10195327689872 (each), 40195327689873 (case), Lot Numbers: 24HMK254

View the original FDA recall notice

Why was this product recalled?

Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods

What actions did MEDLINE INDUSTRIES, LP - Northfield take?

An IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 11/7/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.comRecall Reference #: R-24-209-FGX1 Recall Code: Recall Code 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology